Page 24 Instruction manual Parker X-Flow FM-1407 Rev - 11/2021
LED will flash the number of tens and the red LED the number of units in the mode. The flashes are
called “count-flashes” and have a pattern of 0.5 sec. on, 0.5 sec. off. The control mode numbers can be
found at parameter “control mode”.
Read current control mode (In run mode, Press switch 2x briefly)
LED’s
Time
Indication
Green
Red
amount of count
flashes (0…2)
Off
0 .. 2 sec. maximum
tens in control mode number
off
amount of count
flashes (0..9)
0 .. 2 sec. maximum
units in control mode number
Value zero will be indicated by a period of 1 sec. off (0.5 sec. off + 0.5 sec off).
3.7.2 Change control mode:
For switching between different functions in use of a digital controller several modes are available. More
information about the available control modes can be found at parameter “Control mode”.
Pressing the switch 4x briefly with intervals of max. 1 second in normal running/operation mode will
trigger the instrument to “change” its control mode.
Write current control mode (In run mode, Press switch 4x briefly)
step
action
indication
time
handling
1
Set tens of
setpoint / control
mode number
Green LED flashes
0.1 sec on
0.1 sec off
time-out:
60 sec
Press switch and count green flashes for
tens of control mode number.
Release when desired count is reached.
Count-flashes start
when switch is pressed:
0.5 sec on
0.5 sec off
Counts to max. (2) and then starts at 0
again. When counting fails, keep switch
pressed and restart counting for next
attempt.
2
Set units of
setpoint / control
mode number
Red LED flashes
0.1 sec on
0.1 sec off
time-out:
60 sec
Press switch and count red flashes for units
of control mode number.
Release when desired count is reached.
Count-flashes start
when switch is pressed:
0.5 sec on
0.5 sec off
Counts to max. (9) and then starts at 0
again. When counting failed, keep switch
pressed and restart counting for next
attempt.
Instrument returns to normal running/operation mode.
Changes are valid when they are made within the time-out times.
See parameter ‘Control mode’ (section 3.10.1) for table of values.
See Changing Default Control Mode (section 4.1.6) for behaviour at power-up of the instrument.
Value zero will be indicated by a period of 1 sec. off (0.5 sec. off + 0.5 sec off).
When value zero is wanted, press switch shortly and release it again within 1 sec.
Before each action of flash-counting, the LED’s to be used for counting will flash in a high
frequency. (Pattern: 0.1 sec on, 0.1 sec off). As soon as the switch is pressed down, this LED
(or both LED’s) will be off, and the counting sequence will start.
3.8 LED indications
LED’s
Time
Indication
Green
Red
slow wink
0.2 sec on,
0.2 sec off
Wink mode
By a command send to the instrument.
fast wink
0.1 sec on,
0.1 sec off
Switch-released, selected action started.
Green and Red LED turn-by-turn indication modes (no switch used)