Bulletin 75-00 J / Page 7
V
LIQUID SOL
VENT / WATER SOL
MAIN MENU
START UP
SETUP
HISTORY
OTHER SETTINGS
SELECT WITH V PRESS (ENTER)
PRESS (MENU) TO CANCEL
Main Menu Options
V
LIQUID SOL
VENT / WATER SOL
SETUP
SET TIME AND DATE
PURGE OPTIONS
CALIBRATE PRESURE INPUT
TRIM TEMPERATURE READING
CHANGE LIQUID FEED SETTINGS
SELECT WITH V PRESS (ENTER)
PRESS (MENU) TO CANCEL
Setup Menu
SET TIME/DAY
WHAT IS THE DAY? TUE
SELECT WITH V PRESS (ENTER)
PRESS (MENU) TO CANCEL
V
LIQUID SOL
VENT / WATER SOL
Time / Date Setup Screen
PURGE
DURATION
01
05 MIN
02
05 MIN
SELECT WITH V PRESS (ENTER)
PRESS (MENU) TO CANCEL
V
LIQUID SOL
VENT / WATER SOL
Purge Point Duration Setup Screen
PURGE OPTION
AUTO
MANUAL
TIME BASED
SELECT WITH V PRESS (ENTER)
PRESS (MENU) TO CANCEL
V
LIQUID SOL
VENT / WATER SOL
Purge Option Menu
return to the main menu.
Control Features Setup
Th
is includes date, time, purge type (automatic, manual,
etc.), and the number of purge points being sampled
1. Press and release the “Menu” button on the keypad. Th
e
rapid purger setup main menu will appear.
Use the up/down arrow buttons to select the “Setup” op-
tion. Press and release the “Enter” key.
Press and release the “Enter” key a second time to enter
the current date. Use the up/down arrow keys to select
the correct month. Press and release the “Enter” button
to move to the next fi eld. Repeat these steps to enter the
day and year. When fi nished, press, and release the “En-
ter” key. Enter the current day of the week by again using
the Up/down arrow keys. Next, enter the current time
using the same set of key strokes. Don’t forget to select
between AM and PM in this screen. When you’re done,
press the “Enter” key to return to the main menu.
2. When in the main operating screen press and release the
“Menu” button. Use the up/down arrow buttons to select
the “Setup” option. Press and release the “Enter” button.
Use the up/down arrow keys to select the “Purge Op-
tions” fi eld. Press and release the “Enter” button, “Setup
Points” will be highlighted. Press and release the “Enter”
key on the keypad. Using the up/down arrow keys, select
how many purge points are being used. When done,
press and release the “Enter” key. Now we must enter the
time (in minutes) each purge point is sampled. Th
is set-
ting is less important when the “Automatic” purge mode
is selected. In this mode, sample times will change up or
down based on vent cycle history. With the point and du-
ration highlighted (darkened bar across the fi eld), press
and release the “Enter” key. An arrow will appear to the
left of the time to be changed. Using the up/down arrow
keys, select a time from 5 – 20 minutes in 5 minute incre-
ments. Press and release the “Enter” key when done. Use
the up/down arrow key to select the next point and dura-
tion to be set. Follow the same procedure for each of the
purge points being used. When done, press and release
the “Menu” key to return to the purge options screen.
3. Use the up/down arrow keys to select the “Purge Type”
option and press and release the “Enter” key. 3 Purge Op-
tions are available to you in this screen.
To select “Auto”, press and release the “Enter” key. Th
is
is all that is required to operate in the “Auto” mode (as
defi ned in this manual). Press and release the “Menu”
key to return to the main screen. If you need to terminate
the active point being sampled and move to the next,
press and release the “Terminate” (TERM) key. Termi-
nate Active Point (1-20) should be highlighted. Press and
release the “Enter” key to move to the next purge point.