10
FilterTec Plus Balance
Balances with capacity of 8,100 grams x 0.1 g resolution with all FilterTec Plus models.
Larger balances available upon request. Contact Parker Customer Service.
FilterTec Plus Software
Main menu with five operational modes:
Constant Rate Mode
: Pumps at a user defined rate, and can switch to a user defined
pressure with five, user-definable alarms, in particular, the
Pmaintain
Alarm that
switches from Rate to Pressure control.
Constant Pressure Mode
: Pressure controlled flow rate is varied to maintain a user-
defined pressure by modulating the Feed Rate, with five, user-definable alarms.
Programmable R-P Mode:
Programmable step-by-step variable Rate and Pressure
control, with access to a six-way valve.
Setup Mode
: Selection of user preferences and interface options.
Manual Mode
: Manual control of motor speed and will monitor one pressure, no alarms.
Can change the Rate or Pressure on the fly
. This is a very useful tool in determining
the optimum parameters for your process.
Optimizes DEF Applications:
Dead End, Direct Flow, or Normal Flow Filtration.
Implement Filterability Studies
, in addition to Filter Size Determinations.
Documentation Software for PC
SciDoc Excel Spreadsheet with custom macros and WinWedge interface software for data
compilation. Sent to you ready to use.
Complete process analysis with graphing of data.
Real-time verification and documentation of process parameters.
Includes graphs of:
Feed Rate (FF), Collection Rate (C1) and Pressure (P1) vs. Time
Collection Rate (C1) and Filtrate Quantity (Q1) vs. Time
T/W vs. Time, Vmax
Display, Printout and Excel Abbreviations:
MT
= Military Time, HH:MM:SS
FF
= Feed Rate, ml/min
RT
= Run Time, 00:00:00 at START
CV
= Cumulative Feed Volume, ml
Q1
= Filtrate Quantity collected
C1
= Collection Rate, gr/min
P1
= Pressure at Sensor 1, psi, bar or kpa
CW
= Clockwise Pump Direction
P2
= Pressure at Sensor 2, psi, bar or kpa
CCW
=Counter Clockwise Pump Direction
P3
= Pressure at Sensor 3, psi, bar or kpa
ST
= Pump Status, START, RUN, etc.
AL
= Alarm, e.g. CV Cumulative Volume Alarm
CV
=1, Cumulative Volume Alarm is “OFF”
HP
=4, High Pressure Alarm is “Pmaintain”
RT
=2, Run Time Alarm is “ALERT ONLY”
LP
=1, Low Pressure Alarm is “OFF”
LF
=2, Low Flow Alarm is “ALERT ONLY”
Q1
=3, Filtrate Weight Alarm is “PUMP STOP”
RF
= Response Factor