Parker Hannifin Corporation
14
GB
Technical specifications
Feature
Specification
Product start-up time
Five seconds minimum
Measurement period
Default 30 seconds run time; 15 seconds data logging time
Reporting interval
Once per second
Principle of operation
Laser diode optical detection of actual particulates
International codes range
Up to ISO 22, NAS 12
Calibration
Calibration by recognised online methods confirmed by the
relevant ISO procedures
MTD – via a certified primary ISO 11171 automatic particle
detector using ISO 11943 principles. Particle distribution reporting
to ISO 4406:1999
Recalibration and Servicing
Required every 12 months
Working pressure
350 bar (5076 psi) maximum
Pressures above 2.5 bar require the use of the PRV
(ACC6NN027)
Flow range
icountOS system regulates flow to an optimal 60ml/minute
(0.016 USGPM)
Fluid connection interface
INLET: 6mm push-fit
DRAIN: 4mm push-fit
Ambient storage temperature
for unit
–40ºC to +80ºC; –40ºF to +176ºF
Operating temperature for unit
–30ºC to +60ºC; –22ºF to +140ºF
Operating humidity range
5%RH to 100%RH
Fluid operating temperature
(Oil)
+5ºC to +60ºC; +41ºF to +140ºF
Fluid operating temperature
(Fuel)
–20ºC to +60ºC; –4ºF to +140ºF
Moisture sensor
Linear scale within the range 5%RH to 100%RH
Moisture sensor stability
±2%RH typical at 50% in one year
Moisture sensor calibration
±5%RH (over a compensated temperature range of +10ºC to
+80ºC; +50ºF to +176ºF)
Computer compatibility
Parker supplies an IP68-rated RJ45 connection that may be
connected to a laptop computer’s RJ45 LAN port using the 2m
cable supplied.
Power requirements
Supplied by an internal battery and/or an external power supply.
Parker recommend that the battery is put through a full
discharge/full recharge cycle every three months. Please
contact Parker for advice or if in doubt.
Certification
IP54 rating (unit open)
IP67 rating (unit closed)
EC Declaration of Conformity (see page 6)