• Place all the individual parts in the prepared solution and
leave them to soak.
• Clean the single parts thoroughly with a brush if necessary.
• Thoroughly rinse all parts under hygienically tested water
and leave them to dry completely on a clean, dry and ab-
sorbent surface.
• Dispose of the used solution.
VALIDATION
The method has been validated using:
– cleaning instrument disinfectant Bomix
®
plus 2% (Bode)
– Treatment time: 5 minutes
Info:
If the application period is exceeded significantly, the
plastic parts may take on the smell of the disinfectant.
CAUTION
Inadequate disinfection encourages the growth of bacteria
and thus increases the
risk of infection
. Adequate cleaning
with disinfection can only be assured if the specified mixing
ratio and application time are adhered to, and if all individual
parts are completely immersed in the solution for the entire
application time. There must not be any air pockets or
bubbles.
CAUTION
Disinfectant residues on the product can cause
allergic re-
actions or irritation of the mucous membrane
!
The diluted disinfectant solution can be disposed of down
the drain.
– 63 –