Static electricity
discharge (ESD)
IEC 61000-4-2
/EN 61000-4-2
Contact discharge: ±8kV
Air discharge : ±2kV, ±4kV, ±8kV, ±15kV
Contact discharge: ±8kV
Air discharge : ±2kV, ±4kV, ±8kV, ±15kV
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The PD-200 Series is intended for use in the following electromagnetic environment. Users or customers must use
bed in this environment.
Immunity test
IEC 60601-1-2/EN 60601-1-2
Test level
Compliance level
Electrical fast
transient/burst
IEC 61000-4-4
/EN 61000-4-4
±2kV power supply line
±2kV power supply line
Surge
IEC 61000-4-5
/EN 61000-4-5
Line - line: ±0.5kV, ±1.0kV
Line - ground: ±0.5kV, ±1.0kV, ±2.0kV
Line - line: ±0.5kV, ±1.0kV
Line - ground: ±0.5kV, ±1.0kV, ±2.0kV
Short interuptions
IEC 61000-4-11
/EN 61000-4-11
0%UT: 250/300 cycles
0%UT: 250/300 cycles
Note: UT is the AC power voltage before applying the test level.
Power frequency
magnetic fields
IEC 61000-4-8
/EN 61000-4-8
30 A/m
30 A/m
Voltage Dips
IEC 61000-4-11
/EN 61000-4-11
0%UT: 0.5 cycles
Phase: 0
0
, 45
0
, 90
0
, 135
0
, 180
0
, 225
0
,
270
0
, 315
0
0%UT: 0.5 cycles
Phase: 0
0
, 45
0
, 90
0
, 135
0
, 180
0
, 225
0
,
270
0
, 315
0
0%UT: 1 cycles
70%UT: 25/30 cycles
Single phase: 0
0
0%UT: 1 cycles
70%UT: 25/30 cycles
Single phase: 0
0
47