PGC, Inc.
Appendix B; SmartPad version 0.503
Steady State & Programmable
November 2005
B-26
the display should have moved all the way through the temperature indication and back to the
beginning. If it is entered as desired, depress the "TAB" key.
6.
The first digit of the "RH set point" should be flashing. On the number pad depress 450.
It is not necessary to enter the decimal point. The flashing portion of the display should have
moved all the way through the RH indication and back to the beginning. If it is entered as
desired, depress the "TAB" key.
7.
The first digit of the "segment" should be flashing. Depress the "NEXT" key. This
should advance you to segment 02. NOTE - At this point you may depress any digits on the
number pad to take you to any segment (00 to 59).
8.
Enter segments 02 through 07 as above. Remember a field must be completely filled,
and the flashing portion of the display returned to the beginning for the data you are entering to
be accepted. If you leave a field by pressing any of the function keys before the field is filled,
the previous data will remain in memory.
9.
Press the "INIT" key, the screen will change to the "PROGRAM INITIALIZATION"
screen.
10.
The first digit of the "start at segment" should be flashing. On the number pad depress
01. The flashing portion of the display should have moved all the way through the indication
and back to the beginning. If it is entered as desired, depress the "TAB" key.
11.
The first digit of the "loop from segment" should be flashing. On the number pad depress
07. The flashing portion of the display should have moved all the way through the indication
and back to the beginning. If it is entered as desired, depress the "TAB" key.
12.
The first digit of the "loop to segment" should be flashing. On the number pad depress
02 or 07 as selected. The flashing portion of the display should have moved all the way through
the indication and back to the beginning. If it is entered as desired, you MUST depress the
"EXEC" key. This will input your new loop information, instruct the SmartPad™ to change to
the "METER" screen, and the program will begin running at segment 01. If the time does not
count down, you are in a "halt" condition, depress F1 to cause the program to "run".
VERIFYING THE PROGRAM
1.
You may verify the program at any time. From the "METER" screen press the "PROG"
key, the screen should change to the "PROGRAM" screen.
2.
If the lock out function is set to "none", the first digit of the "segment" should be
flashing. You may verify or modify the data in any segment at this time. Depress the "NEXT"
key. This will advance you to segment 01. NOTE - At this point you may depress any digits on
the number pad to take you to any segment (00 to 59). If the lock out function is set to "all", you
will have to use the "NEXT" key to advance through the program, you will be able to look at the
program but not modify it, you also will not be able to start or stop the program.
Summary of Contents for 9354-4250
Page 2: ...PGC Inc Horizontal 500 1000 CFM June 2006 Parameter Generation and Control Inc 2 ...
Page 42: ...PGC Inc Appendix B SmartPad version 0 503 Steady State Programmable November 2005 B 1 ...
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......