PGC, Inc.
Appendix B; SmartPad version 0.503
Steady State & Programmable
November 2005
B-23
SET POINT ENTRY
(Steady State Mode)
The SmartPad™ has been designed to be as user-friendly as possible. To enter the desired set
point for the chamber, press the “
SP
” key to enter the Set Point Screen. This can be done from
the Process Variable Screen. If the controller is displaying any screen other than the process
variable screen, it will be necessary to press “
ESC
” to return to the process variable screen in
order to access the set point screen. When the SmartPad™ displays the set point screen enter the
new set point, no decimal required.
Note
: Since there is no “
ENTER
” key, it is necessary to fill every digit in the field for proper
entry. For example: 10.0°C would be entered as “
1
”, “
0
”, “
0
”. If the field is not filled, the new
set point will be ignored.
After the new set point has been entered, it is usually a good idea to return the controller to the
process variable screen so that the process variables can be monitored.
NOTE
- If lock out option “ALL” has been chosen from the configuration options screen, there
will be no flashing cursor or “TAB” function, allowing viewing only of the set points.