Chapter 3 Function
3-33
3
Alarm Delay Times
This function holds the previous normal value up to the setting number of
times if an alarm is issued.
The previous normal value will be kept on hold up to the setting number of delay times
if an alarm is issued (where measurement is disabled owing to lack of light).
Alarm output (OUT3) will be turned ON when the alarm status continues more than
the setting number of delay times, and analog output and digital output will be turned
ON according to the setting for analog output at alarm and the setting for digital
output at alarm, respectively.
This function is effective if the user does not require the output of
instantaneous alarms resulting for the surface condition changes of
measurement objects.
Setting range
Function
Initial value
0 to 65534
0 (OFF) to 65534 times
65535
Holds the measurement value
before the alarm.
●
Setting Procedure
1
Display the "Alarm Setting" screen while
in setting mode.
2
Press the [ENTER] key. The "Analog
output at alarm" screen will be displayed.
3
Press the [DOWN] key twice. The "Alarm
Delay Times" screen will be displayed.
4
Press the [ENTER] key. The number of
alarm delay times will be displayed. The
leftmost digit will be selected.
5
Press the [UP]/[DOWN] keys together to
change the selected digit.
6
Press the [UP]/[DOWN] key to change the
value of the selected digit. (Example:
Changing the value to 0 times (OFF).)
7
Press the [ENTER] key and determine the
number of delay times.