TEST AND MEASUREMENT
Issue 1
Section 6
MCUK001002C8
Revision 0
– 52 –
Service Manual
Figure 6.29: Temperature calibration 2
Charging Calibration
1.
On the Interface Box, ensure that the EXT PWR switch is ON. Check that the BCVCU 82 Ohms switch is set to ON.
2.
Holding down the Shift key, press F4. Select PROG CHARG-DAC DATA.
Figure 6.30: Charging calibration 1
3.
Enter a value of 439.
4.
Highlight CNTROL OUT and, using the left / right cursor keys, select CHAR ON. Press ENTER.