Table 9: Available measurement values and parameters
in the composite TB
Composite TB measurements
and parameters
Measure-
ment
Para-
meter
UMPU_SERIAL_NUMBER
R
TOTALIZER_CMD
R/W
SENSOR_SERIAL_NUMBER
R
MULTIK_ACTIVE
R/W
MULTIK_TYPE
R/W
MULTIK_PAIRS
R/W
KVINPUT_SELECTION
R/W
ENABLE_ACTIVE_TW
R/W
CALIBR_MODE_SELECTION
R/W
PATH_ERROR_HANDLING
R/W
UNIT_TYPE_DIMENSION
R/W
UNIT_TYPE_TIME
R/W
UNIT_TYPE_VISCOSITY
R/W
UNIT_TYPE_STD_VOL
R/W
SYSTEM_SERIAL_NUMBER
R
FTPA_SERIAL_NUMBER
R
VOLTAGE_SELECTION
R/W
ATTENUATOR_SELECTION
R/W
D.6.1 Clearing the totalizer
Batch totals can be controlled through foundation fieldbus
(see
figure 29
below). The user can start, stop, or reset batch
totalizers by setting the option on the
TOTALIZER_CMD
function of the
composite transducer block
. To set the
totalizers from the foundation fieldbus:
1. Verify that the gate and ground terminals on the main
PCB are connected.
2. Program the
CALIBR_MODE_SELECTION
parameter
on the
composite transducer block
to
gate input
.
After these steps are complete, you can control the batch
totalizer (start, stop, or reset) by selecting the desired option
on
TOTALIZER_CMD
and writing the changes to the meter.
No password is required for this function.
IMPORTANT:
The inventory totalizer can only be reset at the factory.
Figure 29: TOTALIZER_CMD function on composite TB