Table 7: NAMUR NE107 errors and XMT1000 default categories
Error
Sub-error description
Default category
CRC configuration error
Persistent parameter CRC fault
Failed
Software failure
Stack overflow failure
Failed
Sequence or windowed watch dog failure
Failed
Software fault
Failed
Device initialization failure
Initialization failed
Failed
Hardware failure
ADC Bit test fault
Failed
VGA test fault
Failed
Clock frequency fault
Failed
CPU test fault
Failed
Invariable flash memory fault
Failed
Variable memory fault
Failed
FPGA configuration CRC fault
Failed
Temperature test fault
Failed
Driver failure
Failed
Watch dog failure
Failed
DSP hardware errors
Failed
Default ISR (DSP exception)
Failed
DSP exception
Failed
Modbus communication loss
No Modbus communication
Failed
Sensor value out of range
Velocitywarning
Offspec
Device not calibrated
NotCalibrated
Offspec
Sensor measurement error
SingleChAccuracy
Offspec
MultiChAccuracy
Offspec
ActiveTw
Offspec
CycleSkip
Offspec
Amplitude
Offspec
SignalQuality
Offspec
VelocityRange
Offspec
SoundSpeed
Offspec
SNR
Offspec
Device in configuration mode
In configuration mode - indication
Check
Unsupported parameter warning
Unsupported parameter - warning
Check
Invalid parameter range warning
Invalid parameter range - warning
Check
Invalid request warning
Invalid request - warning
Check
Invalid user warning
Invalid User - warning
Check
57
Summary of Contents for PanaFlow XMT1000
Page 1: ...PanaFlow XMT1000 User s manual 910 313 Rev A...
Page 2: ...ii...
Page 4: ...no content intended for this page iv...
Page 79: ...71...