910-198E
8/19/03
Page 5-24
Model 8500
Note:
The step above will be reversed if the Primary Display is set
to Bottom.
5.9
Alarm
The Alarm feature allows the user to view and change the Low and
High Alarm Reference values (thickness set points that include current
gage units and resolution) via the Setup menu. When this feature is
enabled, the Alarm sounds when any displayed reading (actual,
minimum, or maximum) is either less than or equal to the Low Alarm
Reference Value or greater than or equal to the High Alarm Reference
Value.
A flashing Alarm flag, located in the thickness display area, indicates
the alarm condition and repeats an audible beep. The Alarm is
recorded in the Datalogger second status box for all stored
measurements. An “A” indicates the Alarm Mode, an “L” indicates the
Low Alarm condition, and an “H” indicates the High Alarm condition.
To enter the Alarm Setup screen, press [SETUP]. Use the [ ] and [ ]
keys to select Alarm and press [ENTER]. The gage displays the Alarm
Setup screen: