910-198E
8/19/03
Page 7-6
Model 8500
In order to enable communication between the Magna-Mike 8500 and
other devices, the correct cable must be used and both communication
parameters must be set to match the configuration of the computer. To
verify compatibility proceed as follows:
1.
Press the [SETUP] key.
2.
Use the [ ] and [ ] keys to select Communication, and press
[ENTER].
The current communication parameters will be displayed. Use the
[ ] and [ ] keys to select a parameter, and the [
] or [
] key
to change the parameter. The settings options for each parameter
is as follows:
Baud Rate:
(1200, 2400, 4800, 9600, or 19200)
Number of Stop Bits: (1 or 2)
Parity Bit:
(None, Odd, or Even)
Comm Protocol:
(Single or Multiple)
Serial I/O Format:
(F1, F2, F5, F6, F7,or F8)
Foot Switch:
(Send, Save, and Measure)
Max Key:
(Send, Save, and Measure)
Save/Send:
(Primary and Secondary)
Note:
Make sure the settings in the receiving device match that of
the Magna-Mike 8500:
Baud Rate: (1200, 2400, 4800, 9600, or 19200)
Number of Stop Bits: (1 or 2)
Parity Bit: (None, Odd, or Even)
3.
Press [MEAS] to exit and return to Measure mode with the new
parameters.