DigitalFlow™ XMT868i Startup Guide
13
Chapter 1. Installation
1.7.5.2
Wiring an Alarms Option Card
Each alarms option card includes two or four general-purpose
Form C relays
(designated as A, B, C and D).
The maximum electrical ratings for the relays are listed in Chapter 4,
Specifications
. Each of the alarm relays can be
wired as either
Normally Open
(NO) or
Normally Closed
(NC).
In setting up an alarm relay, it may be wired for either
conventional
or
fail-safe
operation. In fail-safe mode, the
alarm relay is constantly energized, except when it is triggered or a power failure or other interruption occurs. See
Figure 4 for the operation of a normally open alarm relay in both conventional and fail-safe mode.
Connect each alarm relay in accordance with the wiring instructions shown on the label inside the rear cover (see
Figure 3 on page 6 and Figure 11 on page 24).
Figure 4: Conventional and Fail-Safe Operation
C
NO
NC
NC
C
NO
NC
NC
C
NO
C
NO
Conventional
not triggered
not triggered
Fail-Safe
triggered or power failure
Fail-Safe
triggered
Conventional
ALARM
MONITORING
DEVICE
DEVICE
MONITORING
ALARM
DEVICE
MONITORING
ALARM
DEVICE
MONITORING
ALARM
Summary of Contents for DigitalFlow XMT868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 28: ...Chapter 1 Installation 20 DigitalFlow XMT868i Startup Guide...
Page 34: ...Chapter 1 Installation DigitalFlow XMT868i Startup Guide 26...
Page 52: ...Chapter 2 Initial Setup 18 DigitalFlow XMT868i Startup Guide...
Page 76: ...Appendix A CE Mark Compliance 10 DigitalFlow XMT868i Startup Guide...
Page 86: ...Chapter 20 DigitalFlow XMT868i Startup Guide no content intended for this page...
Page 90: ...2 DigitalFlow XMT868i Startup Guide...
Page 92: ...2 Doc Tittle Manual Type...
Page 93: ......