DigitalFlow™ XGS868i Startup Guide
57
Appendix A. CE Mark Compliance
Appendix A. CE Mark Compliance
A.1 Introduction
For CE Mark compliance, the Model XGS868i flow transmitter must be wired in accordance with the instructions in this
appendix.
IMPORTANT:
CE Mark compliance is required only for units intended for use in EC countries.
A.2 Wiring
The Model XGS868 must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to Table 13 for the specific requirements.
Note:
If the Model XGS868 is wired as described in this appendix, the unit will comply with the EMC Directive
2004/108/EC.
Table 13: Wiring Requirements
Connection
Cable Type
Ground Termination
Transducer
Armored RG62 a/U
Grounded via cable gland.
Input/Output Armored 22 AWG shielded (e.g. Baystate #78-1197) with
armored material added to outside of jacket
Grounded via cable gland.
Power
Armored 14 AWG 3 conductor
Grounded via cable gland.
Summary of Contents for DigitalFlow XGS868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 48: ...DigitalFlow XGS868i Startup Guide 40...
Page 60: ...Chapter 3 Operation 52 DigitalFlow XGS868i Startup Guide no content intended for this page...
Page 72: ...Appendix B Data Records 64 DigitalFlow XGS868i Startup Guide no content intended for this page...
Page 78: ...Index 70 DigitalFlow XGS868i Startup Guide no content intended for this page...
Page 80: ...Warranty 72 DigitalFlow XGS868i Startup Guide no content intended for this page...
Page 82: ...74 DigitalFlow XGS868i Startup Guide no content intended for this page...
Page 84: ...76 DigitalFlow XGS868i Startup Guide no content intended for this page...
Page 85: ......