Chapter 1. Programming Site Data
1-25
DigitalFlow™ GC868 Programming Manual (2-Channel)
7.
Press
[F1]-[F4]
to select the desired
Error Response
option, as described in
Table 1-6 below.
8.
Press
[F1]
-
[F4]
to set up another output or press
[EXIT]
to select another slot for
set up.
1.7.3.2.2 Setting Up Alarms
1.
Use the
[F1]-[F3]
keys to set up alarm relays A, B, or C, respectively.
Note:
The set up of alarm A is used as an example. Identical procedures would be
used to set up the other alarms.
2.
Press
[F1] = OFF
to disable Alarm A and return to the previous prompt, or press
[F2] = HIGH, [F3] = LOW or [F4] = FAULT
to specify the type for Alarm A.
3.
Press
[F1] = NO
for standard operation or
[F2] = YES
for
Failsafe
operation.
4.
Use the [
←
]
and
[
→
]
and
[F1]-[F4]
keys to select a
Channel
option for the data
source.
5.
The prompts now vary with your choice in step 2.
• If you selected
FAULT,
proceed to step 6.
• If you selected
HIGH
or
LOW,
complete the steps below.
a.
Use the
[
←
]
and
[
→
]
and
[F1]-[F4]
keys to specify the desired
Output
Measurement parameter.
b.
Enter a value for the trigger point of the alarm and press
[ENT]
.
6.
To select the
Fault Type
that will trigger the alarm, press
[F1] = FLOW, [F2] =
OTHER
(non-flow), or
[F3]= ALL.
7.
Press
[F1]
-
[F3]
to set up another alarm or press
[EXIT]
to select another slot for
set up.
1.7.3.2.3 Setting Up
Totalizer/
Frequency Outputs
1.
Press
[F1]-[F4]
to set up outputs A, B, C or D, respectively.
2.
Press
[F1] = OFF
to disable output A and return to the previous prompt, or press
[F2] = FREQ or [F3]
= TTLZR
to designate output A as a frequency or a totalizer
output, respectively.
Note:
The setup of output A is used as an example. Identical procedures would be
used to set up the other outputs.
3.
Use the [
←
]
and
[
→
]
and
[F1]-[F4]
keys to select a
Channel
option for the data
source.
• To program a frequency output, go to step 4.
• To program a totalizer output, go to step 5.
1.7.3.2.4 Frequency Output
4.
The
FREQ
output,
[F2]
, produces a frequency pulse that is proportional to the
output measurement. Complete the steps below to program the frequency
output.
a.
Use the
[
←
]
and
[
→
]
and
[F1]-[F4]
keys to specify the desired
Output
Measurement parameter.
b.
Enter the
Base
value for the low end of the frequency output range and press
[ENT]
.
c.
Enter the
Full
value for the high end of the frequency output range. Press
[ENT]
.
Table 1-6: Error Response Options
Option
Display Response
HOLD
Holds the last “good” reading.
LOW
Forces the outputs to the low set point.
HIGH
Forces the outputs to the high set point.
OTHER
Allows user to enter 4-20 mA error level in mA.
Summary of Contents for DigitalFlow GC868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Warranty iv DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Page 10: ...Contents viii DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Page 130: ...Index 4 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...