EmStat3-MUX8 for OEM 

3.3

 

MUX8 configurations 

 

The multiplexer can be used in different configurations or modes. Each configuration is set by a 
number of jumpers which are located on the board as shown in the illustration of the board layout 
below. 
 
 

 

 
Possible sensor configurations are: 

 

Sensor arrays with (up to) eight working, reference and counter electrodes  

 

Sensor arrays with eight working and eight combined reference/counter electrodes  

 

Sensor arrays with eight working electrodes sharing a reference and a counter electrode  

 

Sensor arrays with eight working electrodes sharing a combined reference/counter 
electrode  

In all configurations the sensors can be multiplexed, leaving the not-selected sensors or cells at 
open circuit. 
Sensor configurations 2, 3 and 4 have the possibility to leave not-selected sensors or cells at open 
circuit or to apply the same potential to all sensors or cells. 

 

Jumpers: 

 

J1: RE to CE 

 

Is placed when the sensor has a combined reference and counter electrode. This jumper therefore 
connects RE to CE. 

J2: CE Direct 

 

If the sensor array has more than one working electrode, but one counter and/or reference 
electrode, this jumper is placed. CE from PalmSens is connected directly to pin 1 and pin 2 of CON1 

J3: RE Direct 

 

If the sensor array has more than one working electrode, but one counter and/or reference, this 
jumper is placed. RE from PalmSens is connected directly to pin 13 and 14 of CON1 

J4: CE MUX 

 

This jumper is placed if CE has to be multiplexed. This is the case when each of the sensors has its 
own counter electrode using pin3 to 10. 

J5: RE MUX 

 

This jumper is placed if RE has to be multiplexed. This is the case when each of the sensors has its 
own reference electrode using pin 15 to 22 

J6: Unselected WE to AGND:    

If placed, all unselected working electrodes or sensors will remain polarized at the potential set by 
the PalmSens or EmStat. If this jumper is not placed only the selected channel is polarized leaving 
the not-selected WE’s at open circuit. 
When using Conf 1 and CE and RE are multiplexed (J4 and J5 are placed), this jumper is not 
relevant since only the selected channel’s WE, CE and RE are polarized. 

J7: Unselected WE to standby potential   

 

See section 3.3 

 

Summary of Contents for EmStat3-MUX8

Page 1: ...EmStat3 MUX8 for OEM Connecting a MUX8 board to an EmStat3 Last revision May 27 2015 2015 PalmSens BV www palmsens com...

Page 2: ...description 4 2 3 Pinout of sensor connector CON4 4 2 4 CON1 to MUX8 5 3 MUX8 PCB 5 3 1 Functional diagram 5 3 2 Set standby potential with the Vaux pin 6 3 3 MUX8 configurations 7 3 4 Electrode conne...

Page 3: ...nd peripheral control 2 Connecting EmStat board to MUX8 multiplexer EmStat3 PCB layout The board has two pin headers CON1 and CON2 CON1 is used to connect for instance the MUX multiplexer or any other...

Page 4: ...x 20 mA when powered from USB 8 5 V analog max 10 mA 9 AGND 10 WE 11 RE 12 CE 2 2 CON2 description Pin Function 1 Tx serial port 2 Rx serial port 3 Reserved DO NOT CONNECT 4 Voltage reference 4 096 V...

Page 5: ...Stat is to be connected to the MUX board using CON1 This can be done in two ways Two ways of connecting the EmStat PCB to the MUX8 PCB 3 MUX8 PCB The MUX8 multiplexer is meant for use with 2 or 3 elec...

Page 6: ...to standby potential PLACED When using Conf 1 and CE and RE are multiplexed J4 and J5 are placed the Vaux will not be applied on the unselected channels since only the selected channel s WE CE and RE...

Page 7: ...his jumper therefore connects RE to CE J2 CE Direct If the sensor array has more than one working electrode but one counter and or reference electrode this jumper is placed CE from PalmSens is connect...

Page 8: ...nter electrodes Jumpers placed are J1 J2 and J3 Note the combined reference counter electrodes are connected to the leads CE Direct and or RE Direct When J6 is also placed the potential is applied to...

Page 9: ...ounter electrode Sensor arrays with eight working electrodes all sharing the same combined reference counter electrode On resistance WE CE and RE channels 2 ohm typical see note Leakage current 10 pA...

Page 10: ...10 EmStat3 MUX8 for OEM EmStat3 MUX8 for OEM 4 PCB dimensions 4 1 MUX8 PCB...

Page 11: ...EmStat3 MUX8 for OEM 11 EmStat3 MUX8 for OEM 4 2 EmStat3 PCB...

Page 12: ...12 EmStat3 MUX8 for OEM EmStat3 MUX8 for OEM 4 3 PCB s connected option 1 variable 18 if mini USB is used 25...

Page 13: ...EmStat3 MUX8 for OEM 13 EmStat3 MUX8 for OEM 4 4 PCB s connected option 2 101 30 variable 16...

Reviews: