24
PALM
®
MicroBeam User Manual - 0202 Z
Trouble Shooting List
No video image
8.2 No video image
8.3 Shift between drawing and cutting
check image in oculars
no image:
check illumination of microscope
check position of main-switch on camera's power supply (if applicable)
check connection between camera and camera power supply (if applicable)
check connection between camera and computer; S-VHS-plug must be plugged in framegrabber board
check position of switching mirror in microscope.
check alignment of camera;
the camera must be adjusted parallel to the stage movement, test this by moving the stage with the
arrow keys (in a slow speed)g
check position of laser marker
Calibrate new as described in the User's Manual;
make sure that you are using the settings for the active lens
check LED's on encoder of moving stage:
LED is green:
o.k.
LED is orange or red:
please contact P.A.L.M. Microlaser
Technologies AG