5. Install the Allegro 3D biocontainer into the base of the tote and ensure biocontainer is as level
as possible to the base of the tote. When laying flat, the ports of the biocontainer should align
with the tote grooves (Figure 21).
Note:
The Allegro biocontainer label is located on the top of the system and needs to be facing
upward towards the user.
Figure 21
Ports aligned with grooves in base of the tote
6. Close all clamps on the outlet lines and ensure that the inlet lines are open on the biocontainer.
Secure the lower swivel door and replace locking pin. Attach the four hoist clips to the holes
four holes located at each corner of the top of the biocontainer. Ensure each clip fastens to
the corresponding hole so that the crossbar is holding the four connection points (Figure 22).
Figure 22
Four hoist clips connected to biocontainer holes
4.4
Filling the Allegro biocontainer/system
Once the biocontainer is secured into the base of the tote and aligned with four hoist connection
perform the following:
1. Check that the outlet lines are closed
2. Using the friction wheel, raise the biocontainer (Figure 23)
Figure 23
Raising the Allegro biocontainer using the friction wheel