Please observe:
The settings for the frequency and the impulse bandwidth must be adjusted in normal operation, i.e.
before pressing the
SETUP
-button for the fi rst time. The intensity of the current can still be varied in the
SETUP-function with the aid of the turn-knob.
ATTENTION:
Having selected a too high initial intensity of the stimulation current may lead to painful stimulation of
the patient during subsequent introduction of the cannula.
Initial parameters for stimulation using the cannula
Cannula-parameter
Adjustment range
Manufacturer’s settings
Stimulation current intensity
0 – 6,00 mA
1,50 mA
Stimulation frequency
1 Hz, 2 Hz
2 Hz
Stimulation impulse bandwidth
50 µs/100 µs/200 µs/300 µs/
500 µs/1,00 ms
100 µs
5.1.3 SETUP LEVEL 3: Setting of the initial parameters for “Catheter mode”
A triple-click of the
SETUP
-button will activate a program option which allows you to individually
preset the stimulation parameters for the catheter after the device is switched on. The parameters and
the manufacturer’s settings are listed below.
Depressing the
SETUP
-button for a longer period of time (min. 2 seconds) will permanently store the
values currently indicated on the display. A monitoring sound is audible during the store-procedure,
then you are redirected to the PAUSE-mode. A short depression of the
SETUP
-button will guide you to
the next item of the SETUP-menu without having changed the existing settings (see 3.8.4).
Please observe:
The settings for the frequency and the impulse bandwidth must be adjusted in normal operation. i.e.
before pressing the SETUP-button for the fi rst time. The intensity of the current can still be varied in the
SETUP-function with the aid of the turn-knob.
ATTENTION:
Having selected a too high initial intensity of the stimulation current may lead to painful stimulation of
the patient during subsequent introduction of the cannula.
Initial parameters for stimulation using the catheter
Catheter-parameter
Adjustment range
Manufacturer’s settings
Stimulation current intensity
0 – 20,00 mA
2,00 mA
Stimulation frequency
1 Hz, 2 Hz
1 Hz
Stimulation impulse band-
width
50 µs/100 µs/200 µs/300 µs/500 µs/
1,00 ms
0,5ms
5.1.4 SETUP LEVEL 4: Resistance indication with warning sound notifi cation
Pressing the SETUP-button briefl y four times will activate a program option which allows you to adjust
the signal sound for the patient resistance.
Here, the threshold value at which the acoustic signal for the patient resistance will change can be
adjusted. Single sound pulses are emitted below this value (one single sound pulse for every patient
stimulation impulse). If the resistance of the patient circuit is higher than the adjusted threshold, this is
signalized acoustically with two short sound pulses for every patient stimulation impulse. The function
is deactivated with the setting value 0. This setting applies for cannula and for CATH.
V
0
1
2
3
4
5
Th
re
de
5.
AT
Pr
th
of
of
In
w
V
0
1
5.
Bi
M
th
sit
to
pr
ra