background image

English

27

he 

ly 
l-

r-

-

-

-

o-

a)

es 

-

re 

Dynamic electrical- and dynamic magnetic interference fi elds may interact with the device and the system. 
These interactions may have an infl uence on the measurement of the actual stimulus current, and in the 
extreme case, they may lead to an error in indication and perhaps to a safety-shutdown of the device.

The MultiStim SWITCH may not be used near equipment emitting strong magnetic fi elds, e.g. radio-
telephones, surgical RF-devices, short wave- or microwave-therapy devices. Any potential introduction 
of high-frequency currents into the stimulation cannula may lead to a damaging of the nerves. 
The device may not be connected with other equipment. If the MultiStim SWITCH is operated nearby 
to another device, then the devices or the system must be monitored and the correct functioning as 
intended must be verifi ed in this arrangement as it is being used. 

Under unfavorable conditions, the MultiStim SWITCH may disturb other equipment in its function. We 
therefore recommend to check all other equipment and facilities for compatibility with the MultiStim 
SWITCH, and if necessary, to remove these from the patient. 

The operation of other equipment or systems with the accessories may lead to increased emissions or 
to reduced interference immunity of these systems. Please observe the enclosed EMC (electromagnetic 
compatibility) information regarding installation, start-up and operation of the equipment or system 
(see chapter Electromagnetic Compatibility (EMC)). 

To avoid damaging the connector cable and the device, please do not hold or carry the device by its con-
nector cables or by its accessories. Do not wind the cable around the device or around other facilities. 

The winding of the connector cable during normal operation of the stimulator will produce inductive 
components and may, if very short stimulation pulses are being used, lead to a reduction of stimula-
tion performance or to faulty measurement of the actual stimulation current. Misinterpretations of the 
reported values can be the consequence.

In case of battery leakage, the device should not be operated anymore for security reasons the device 
must be returned to the manufacturer for proper cleaning. 

Avoid unclean connections. Water and dirt will deteriorate the contact properties of the plug-in connec-
tions and lead to unintentional short-circuits or leakage-currents. These may lead to partial- or even total 
diversion of the stimulation currents, which will cause the stimulation effect to decline, or even to fail 
completely. In this case, the device cannot display the actually fl owing patient current correctly any more. 

To avoid damaging the MultiStim SWITCH and its accessories, do not use aggressive cleaning agents, 
further details may be found under item “Cleaning and Disinfection of the Device”. Check all accesso-
ries in regular intervals. The insulation of the lines and hardware connections may show no damages. 

The User Instructions are to be followed for the operation of the MultiStim SWITCH and the corre-
sponding accessories. When sterile accessories are being used, please always provide for sterile environ-
mental conditions.

Avoid unintentional bone contact with the stimulation cannula, since the cannula may thereby be 
substantially damaged which may consequently cause a traumatization of the tissue.

Keep the accessories and MultiStim SWITCH away from live / current bearing objects. The electrostatic 
and electromagnetic fi elds radiated therefrom may have an infl uence on the stimulation result, and 
these themselves may even lead to unwanted stimulation effects in the tissue.

The device, the connector cables and the plugs must be kept completely clean and dry before and 
during the application. Moisture and uncleanness will affect the function of the nerve stimulator and/or 
the stimulation result. 

Please note the position of metal implants in the tissue (e.g. plates or electrode cables). They may 
perhaps conduct the stimulation signals to other locations and can cause detrimental effects there. 
Implanted electronic equipment may be impaired by the stimulation current, which in turn will lead to 
a malfunctioning of these implants or may even destroy them. 

To avoid malfunctions of the MultiStim SWITCH, please check all functions before the intervention and 
make sure that the accessories are suitable for the application. The accessories used must correspond to 
safety class type BF. 

To protect the patient from electrophysiological shock by electrostatic discharge (ESD), it is necessary to 
wear corresponding clothing and to move about in an appropriately protected environment. An elec-
trostatic discharge (ESD) can cause extremely high current densities to appear at the tip of the cannula, 
which can damage the surrounding tissue.

XS190174B MultiStim SWITCH GB.indd   27

XS190174B MultiStim SWITCH GB.indd   27

12.03.10   13:53

12.03.10   13:53

Summary of Contents for MultiStim SWITCH

Page 1: ...MultiStim SWITCH and MultiStim SENSOR Setting the trend in nerve stimulation Plexus and Epidural Anaesthesia...

Page 2: ...e following advantages The MultiStim SENSOR supports patient friendly percutaneous localization and identification of nerves with the aid of a stimulation handle the PEG electrode Percutaneous Electro...

Page 3: ...ent which can be corrected immediately Integrated safety functions High precision microprocessor controlled adjustment of constant current Device type BF Battery 9 V Current intensity max 6 mApp 20 mA...

Page 4: ...ed through the skin without requiring a puncture which will evoke a reflexive response when the nerve is encountered The cannula is introduced at the insertion point identified by means of this method...

Page 5: ...tatus indicator in the display The intensity and the frequency of the stimulation current as well as the pulse width can be adjusted separately for both outputs This button will remain without functio...

Page 6: ...N OFF PAUSE Pulse width Turning knob Pulse width Battery condition Patient resistance Current intensity Actually flowing current NOMINAL ACTUAL current Active output for example Cannula MultiStim SWIT...

Page 7: ...Y patient cable point PAJUNK has set new standards in electrical nerve stimulation because for the first time ever it provides the option for alternative stimulation by means of cannula or by stimulat...

Page 8: ...mmediate identification and correction of a misplaced cannula before mechanical or chemical injuries The revolution in the nerve stimulation More safety through the indication of pat are caused Becaus...

Page 9: ...o analogous bar graphs The measurement of the resistance connected therewith is only possible if the highly precise PAJUNK stimulation cannula are used An individual threshold value can be predefined...

Page 10: ...be increased to 20 mA and will therefore correspond with the specific requirements of epidural stimulation The intensity and the frequency of the stimulation current as well as the pulse width can be...

Page 11: ...s these steps will be correspon dingly large at great distances to the nerve and will become smaller as the cannula approaches the nerve see accompanying table Automatic adjustment of the current inte...

Page 12: ...n of stimulation cannula and monopolar handle 1151 94 32 Accessories 1 Patient cable for the connection of stimulation cannula 1151 94 13 2 PEG cable for the connection of stimulation cannula and mono...

Page 13: ...Peripheral Epidural Spinal MultiStim SWITCH XS190174B MultiStim SWITCH D indd 1 XS190174B MultiStim SWITCH D indd 1 08 04 10 11 47 08 04 10 11 47...

Page 14: ...TUP functions 35 5 1 1 SETUP LEVEL 1 Volume control 35 5 1 2 SETUP LEVEL 2 Setting of the initial parameters for Cannula mode 36 5 1 3 SETUP LEVEL 3 Setting of the initial parameters for Catheter mode...

Page 15: ...s Suitcase for storing SWITCH and accessories Please observe by all means The patient main cable of the MultiStim SWITCH is compatible with that of the MultiStim SENSOR due to its technical design Ple...

Page 16: ...esive electrode will lead to a malpositioning of the stimulation cannula please ensure that the adhesive electrode which functions as a neutral electrode here is in sufficiently sound contact with a l...

Page 17: ...unintentional short circuits or leakage currents These may lead to partial or even total diversion of the stimulation currents which will cause the stimulation effect to decline or even to fail compl...

Page 18: ...indicated exactly on the display Current constant devices which permit the selection of the current intensity mA for the stimulus impulse have gained acceptance within the last few years For the Multi...

Page 19: ...circuit is closed i e stimulation current is running through the patient The control sound will become more intensive in this case Due to a high stimulus voltage of max 95 Vss this renders an exceptio...

Page 20: ...e impulse bandwidth Mode Impulse frequency 3 4 The elements of the display Symbol Meaning Indicates must current is displayed in mA Indication of the electrical patient circuit Electric circuit not cl...

Page 21: ...0 1 description see table in chapter Setup Level 5 4 Operation 4 1 Pre operational check Please observe Equipment with divergent behavior may not be put into operation In this case please contact the...

Page 22: ...m the one described in 3 and 4 then a faulty cable could be the cause short circuit interruption Check the function once more with a second cable Finally remove the red short cir cuit plug from the so...

Page 23: ...bandwidth are separately adjustable for both of the two outputs see SETUP functions 4 6 Mode catheter CATH Aside of stimulation using an invasive stimulation cannula MultiStim SWITCH addition ally of...

Page 24: ...Incremental step 0 00 mA 0 50 mA 0 02 mA 0 mA 2 mA 0 1 mA 0 50 mA 1 00 mA 0 05 mA 2 mA 5 mA 0 2 mA 1 00 mA 2 00 mA 0 10 mA 5 mA 10 mA 0 5 mA 2 00 mA 4 00 mA 0 20 mA 10 mA 20 mA 1 0 mA 4 00 mA 6 00 mA...

Page 25: ...TUP button for a longer period of time min 2 seconds will permanently store the volume value currently indicated on the display A monitoring sound is audible during the store procedure The device will...

Page 26: ...3 8 4 Please observe The settings for the frequency and the impulse bandwidth must be adjusted in normal operation i e before pressing the SETUP button for the first time The intensity of the current...

Page 27: ...n which allows you to adjust the pulse width for the cannula mode Here the pulse width selection mode can be selected by means of the turnable key button You can either choose manual selection Selecti...

Page 28: ...rned to the manufacturer for repair Please ask your dealer if you need assistance he will be glad to help you Please indicate the error code in the repair order The following error codes are possible...

Page 29: ...of the functionality test before the first operational use of the device 2 Date and time of the familiarization as well as the names of the persons familiarized with the opera tion of the device 3 Dat...

Page 30: ...ding 15 s are possible particularly for fall time 12 Notes warnings If you have adjusted a stimulation current intensity 0 06 mA or 0 2 mA during operation in the CATH mode and 0 00 mA are indicated i...

Page 31: ...laration Electromagnetic Emissions according to EN 60601 1 2 2007 5 2 2 1 Table 1 MultiStim SWITCH is intended for use in the electromagnetic environment specified below The customer or user of the Mu...

Page 32: ...EC 61000 4 4 2 kV for power supply lines 1 kV for input output lines Not applicable Not applicable Surges according to IEC 61000 4 5 1kV line s to line s 2kV line s to earth Not applicable Not applica...

Page 33: ...quation applicable to frequency of the transmitter Recommended separa tion distance d 3 5 U1 P for 150kHz to 80MHz d 12 U1 P for 80MHz to 800MHz d 23 U1 P for 800MHz to 2 5GHz where P is the maxi mum...

Page 34: ...e guarantees to manufacture its products with greatest possible care THIS IS THE ONLY VALID GUARANTEE AND IT SHALL REPLACE ALL OTHER WARRANTIES GIVEN AND REPRESENTATIONS MADE It shall be observed that...

Page 35: ...hthalates as identified Does not contain latex Contains latex Non pyrogenic Pieces Catalog number Unique identifier Batch Identification Use by date Date of manufacture Device Type BF Catalogue Excerp...

Page 36: ...H Medizintechnologie Karl Hall Stra e 1 D 78187 Geisingen Germany Telefon 07704 9291 0 Telefax 07704 9291 600 www pajunk com 0124 XS190174B MultiStim SWITCH E indd 136 XS190174B MultiStim SWITCH E ind...

Reviews: