The user can use every commercially available CE-labeled self-adhesive electrode for medical purposes.
Do not use the product in case of commun incompatibilities and/or interactions of the material.
The increased attention of the user is required at effective current densities beyond 2 mA/cm
2
for all
adhesive electrode surfaces.
A theoretical maximum current density of 32.4 mA/mm
2
can be reached if the stimulation cannula is
used!
Only perform wiping disinfection, no spraying disinfection under all circumstances! Avoid condensa-
tion!
13 Electromagnetic compatibility (EMC)
MultiStim SWITCH complies with the standard for electromagnetic compatibility (EMC) EN 60601-1-
2:2007.
The tests for electromagnetic compatibility were performed by: Nemko GmbH & Co. KG – Testing-
and Certifi cation Authority – Reetzstrasse 58 - 76327 Pfi nztal - Germany
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
(according to EN 60601-1-2:2007; 5.2.2.1 Table 1)
MultiStim SWITCH is intended for use in the electromagnetic environment specifi ed below. The
customer or user of the MultiStim SWITCH should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
- guidance
RF emission
according to CISPR 11
Group 1
MultiStim SWITCH uses RF
energy only for its internal func-
tion. Therefore, its RF emissions
are very low and are not likely
to cause any interference in
nearby electronic equipment.
RF emission
according to CISPR 11
Class B
The MultiStim SWITCH is suita-
ble for use in all establishments,
including domestic establish-
ments and those directly
connected to the public low-
voltage power supply network
that supplies buildings used for
domestic purposes.
Harmonic emissions
according to IEC 61000-3-2
Not applicable
Voltage fl uctuations/fl icker
emissions according to
IEC 61000-3-3
Not applicable