4.5.2.1 TIME BASED TRANSIENT EXAMPLE (cont)
Press
1, ENTER, 1, ENTER, 1,
to select Waveform 1 on all three phases.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=80 Vb=80 Vc=80
F=400 WFa=1 WFb=1 WFc=1
Press
ENTER
, now we are starting segment 2 of 2.
TRANSIENT: PRGM #4 SEGMENT #2 OF 2
T=0.0002 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
ENTER, ENTER, .01, ENTER,
to select default segment 2 of 2 and a time 10 milliseconds.
TRANSIENT: PRGM #4 SEGMENT #2 OF 2
T=.01 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
80, ENTER, 80, ENTER, 80, ENTER,
to select 80 volts on all three phases.
TRANSIENT: PRGM #4 SEGMENT #2 OF 2
T=0.01 Va=80 Vb=80 Vc=80
F=400 WFa=1 WFb=1 WFc=1
Press
400, ENTER,
to select 400 Hz.
TRANSIENT: PRGM #4 SEGMENT #2 OF 2
T=0.01 Va=80 Vb=80 Vc=80
F=400 WFa=1 WFb=1 WFc=1