OMEGA SRT
-
15/150 Dual Superficial /
Rockwell Hardness Tester
4.0 Operation
1.
Mount the sample so it is flat on the stage. For non
-
parallel mounts a leveling vice is
recommended (optional)
-
Sample should be clean
-
Minimum thickness should be 10X the depth of the indentation
-
Do not test samples in a metallographic mounts
-
Sample must be secured so it does not move during loading, otherwise the indenter can
be damaged
2.
Turn on the OMEGA SRT
-
15/150
3.
Select appropriate load and indenter (see following table).