16
Version1.0 rev 18 Feb 2015
Rotr
4. Re-install the retaining ring using a Thorlabs SPW602 filter tool. Be careful not to over-tighten once
the ring is clamping the filter into the installed position.
5. Ensure the filter is fully retained and has no free play by lightly shaking the filter wheel.
6. Re-install the access panel using a flat head screwdriver or similar tool, and screwing clockwise.