15
Version1.0 rev 18 Feb 2015
Rotr
SECTION 6. MAINTENANCE
PLEASE NOTE the following:
•
If the equipment is used in a manner not specified by Andor, the protection provided by the
equipment may be impaired.
•
In case of emergency, the disconnecting device is the mains lead. This will either be the mains lead
connected to the product, or in the case of a cabinet-based system the mains lead to the cabinet.
See Section 5.1 Emergency Disconnection.
6.1. C
LEANING AND
D
ECONTAMINATION
•
To clean the product, only use a damp lint-free cloth. Do not wet the connectors. Do not use
solvents, cleaning agents, or aerosols.
6.2. R
EGULAR
C
HECKS
•
The state of the product should be checked regularly, especially the integrity of the enclosure and
the mains cable.
•
Do not use equipment that is damaged.
6.3. A
NNUAL
E
LECTRICAL
C
HECKS
•
It is advisable to check the integrity of the insulation and protective earth of the product on an
annual basis, e.g. U.K. PAT testing.
•
Do not use equipment that is damaged.
6.4. I
NSTALLING
F
ILTERS
1. Access to the filters is through the access port as indicated. Using a flat head screwdriver or similar
tool unscrew the cover anti-clockwise.
2. Remove the retaining ring using a Thorlabs SPW602 filter tool.
3. Carefully insert the filter, noting the orientation of the filter to suit the image direction (normally arrow
toward camera side).