OWON HDS-N Series Handheld DSO & DMM 10-Appendix
Record length
6K points on each channel
z
5ns/div
~
100s/div, stepping in the "1-2.5-5"
mode----HDS1022M-N
z
5ns/div
~
100s/div, stepping in the "1-2-5"
mode----HDS2062M-N
,
HDS3102M-N
Scanning speed range
(
S/div
)
Sampling rate and delay time
accuracy
±
100ppm
(any time interval which is equal to or larger than 1ms)
Single:
±
(1 sampling interval time +100ppm
×
0.6ns)
>average 16 :
±
(1 sampling interval time +100ppm
×
0.4ns)
Time interval (
△
T)measurement
accuracy
(
full bandwidth
)
Vertical
With the resolution of 8 bits, make sampling on both
channels synchronously.
Analog digital converter
(
A/D
)
z
±
10 div----HDS1022M-N
z
±
1V(5mV/div ~ 100mV/div),
±
40V(200mV/div ~5
V /div) ----HDS2062M-N
,
HDS3102M-N
Displacement range
20 MHz ----HDS1022M-N
60 MHz ----HDS2062M-N
100MHz----HDS3102M-N
Analog bandwidth
Single bandwidth
Full bandwidth
Low frequency response
(
AC coupling, -3dB
)
≥
5Hz
(
at the BNC
)
≤
17.5ns----HDS1022M-N
≤
5.8 ns----HDS2062M-N
≤
3.5 ns----HDS3102M-N
Rise time
(
typical one at the BNC
)
DC gain accuracy
±
3%
The voltage difference (
△
V) between any two points on
the waveform after averaging the captured waveforms
more than 16:
±
(5% r 0.05 divisions).
DC measurement accuracy
(
average
value sampling mode
)
Trigger
Triggering level range
±
6 divisions from the screen center
Triggering level accuracy (typical)
which is applicable to the signal
with rise and fall time equal to or
longer than 20ns
±
0.3 divisions
655 divisions for pre-triggering and 4 divisions for post-
triggering.
Trigger displacement
Trigger Holdoff range
100ns~10s
78