background image

OWON HDS-N Series Handheld DSO & DMM                   6-Using the Oscilloscope 

 

Figure 17: AC-Coupling 

6.16 Reversing the Polarity of the Displayed Waveform   

To invert the input CH1 waveform, do the following:   

1. Press 

the 

MENU

 key and the function menu appears at the right side of the screen. 

2. Press 

the 

MENU

 

 or 

MENU

 

 key to select 

CH1 SETUP

. Four items selectable are 

displayed at the bottom of the screen. 

3. Press 

F4 

key to jump to 

Inverted, 

set it as 

ON

.

 

The inverted waveform of CH1 is displayed 

on the screen. 

Now, you can see a screen that looks like the following figure 18. 

 

Figure 18: Inverted On 

25 

Summary of Contents for HDS-N series

Page 1: ...HDS N Series Handheld Digital Storage Oscilloscope Multimeter User Manual HDS1022M N HDS2062M N HDS3102M N WWW OWON COM HK ...

Page 2: ... However OWON will continue to improve products and reserves the rights to changes specification at any time without notice OWON is the registered trademark of the Lilliput Company Headquarter Fujian Lilliput Optoelectronics Technology Co Ltd The mansion of optoelectronics 19 Heming Road Lantian industrial zone Zhangzhou Fujian China Tel 86 596 2130430 Fax 86 596 2109272 Web www owon com hk Mail B...

Page 3: ... purchase This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care Lilliput shall not be obligated to furnish service under this warranty a to repair damage resulting from attempts by personnel other than Lilliput representatives to install repair or service the product b to repair damage resulting from improper use or conn...

Page 4: ...t Connections 8 5 1 1 Input connections 8 5 1 2 The connection of 1KHz 5V Square wave test signal 9 5 2 Front Panel and Keys Overview 9 6 USING THE OSCILLOSCOPE 12 6 1 About this Chapter 12 6 2 Powering Up the Oscilloscope 12 6 3 Oscilloscope Operation Window 12 6 4 Menu Description 14 6 5 Manually Setting the Vertical System Horizontal System and Trigger Position 15 6 5 1 Setting the Vertical Sys...

Page 5: ...easuring DC Voltage 32 7 4 6 Measuring AC Voltage 33 7 4 7 Measuring DC Current 34 7 4 8 Measuring AC Current 35 7 5 Freezing the Readings 36 7 6 Taking a Relative Measurement 37 7 7 Selecting Automatic ManualRange Adjustment 38 8 ADVANCED FUNCTION OF OSCILLOSCOPE 39 8 1 About this Chapter 39 8 2 Setting the Vertical CH1 and CH2 39 8 2 1 Setting the Channel Coupling 40 8 2 2 Open and Close Setting...

Page 6: ...ormal mode 57 8 11 2 Setting the cursor measurement in FFT mode 60 8 12 Autoscale 61 8 13 Record 63 8 14 Using FFT 66 8 15 System State Menu 72 8 15 1 Real time clock 73 8 16 Setting of Time Base Mode 73 8 17 Data Transmission 75 9 TROUBLE SHOOTING 76 10 APPENDIX 77 10 1 Appendix A Specifications 77 10 1 1 Oscilloscope 77 10 1 2 Meter 79 10 1 3 General Specifications 81 10 2 Appendix B Maintenance...

Page 7: ...uipment for Measurement Control and Laboratory use part 1 General Requirements EN61326 1 2006 Electrical Equipment for Measurement Control and Laboratory use EMC Requirements part 1 General Requirements EN61000 3 2 2000 A2 2005 Electromagnetic Compatibility EMC part 3 Limits Section 2 Limits for Harmonic Current Emissions Equipment Input Current less than equal to 16A per phase EN61000 3 3 1995 A1...

Page 8: ... Optional 1 Oscilloscope and Battery 2 AC DC adapter Oscilloscope Probe 2 grey 3 Multimeter test lead 2 black and red 4 5 USB Mass Storage connect cable 6 Extension module for small capacitance measurement 7 Probe adjustment tools 8 USB communication cable or RS 232C communication cable 9 User Manual 10 CD ROM software 11 Hard carrying case 12 Soft carrying case 13 the output terminal of 1KHz 5V s...

Page 9: ... Ground Earth gr Term ound inal 2 1 2 Safety Terms The following terms may appear on the instrument Danger The term Danger is used in this manual to indicate that when you read this mark personal injury may be caused to you immediately Warning The term Warning is used in this manual to indicate that when you read this mark personal injury may not be caused to you immediately but you need to be cau...

Page 10: ...ion Ground Clip Signal Input Electrical Outlet Probe Oscilloscope AC Adapter The diagram of the ground wire connection when the battery powered oscilloscope is connected to the AC powered PC through the ports Ground Clip Signal Input Oscilloscope Battery power PC Electrical Outlet Probe USB COM Cable It is not allowed to measure AC power when the oscilloscope is AC powered through the adapter or w...

Page 11: ...d technical personnel are permitted to perform maintenance Pay attention to the nominal values of all terminals To avoid fire or electric shock please keep a watchful eye on all nominal values and marks specified for this product Before any connection performed on this product carefully read the user s manual of the product for further information of nominal values No operation is allowed without ...

Page 12: ...and high contrast with adjustable back light Storage and call out of waveforms Automatic setting function provided capable of fast setting Multiple waveform calculation function Implementation of detecting the average and peak values of the waveform Edge video and alternate triggering function RS232 or USB communication ports Multiple Language User Interface Multimeter 3 3 4 digits Volts Amps Ohms...

Page 13: ...ories The accessory list has been described in the picture of Digital Oscilloscope Parts of this manual You can make a check and find whether there is any accessory loss with reference to the Appendix In case of any accessory loss or damage consult the OWON dealer responsible for such a business or the local office of OWON 4 1 3 Make a Check on the Complete Instrument If the instrument is damaged ...

Page 14: ...harging 2 Multimeter test lead 3 Multimeter input jacks including four circular banana jacks The first jack measuring the current 2 A 10A the second jack on the current Measurement the third COM ground input and the fourth measuring voltage resistance Capacitance input 4 Oscilloscope probes 5 Oscilloscope channel inputs the upper one is for Channel 1 CH1 while the lower one is for Channel 2 CH2 6 ...

Page 15: ...1KHz 5V Square wave test signal At the left side of the oscilloscope it is a port for testing 1 KHz 5V square wave signal which is used to adjust the probe shown as Fig 3 Fig 3 The connection of Square wave test signal 5 2 Front Panel and Keys Overview See the following Figure 4 9 ...

Page 16: ...N Series Handheld DSO DMM 5 Input Connections Figure 4 Front Panel Description 1 AC adapter Port 2 RS 232C Port 3 USB Port 4 USB Mass Storage Port 5 Power switch 6 F1 F5 Switch or Adjust options for each menu 10 ...

Page 17: ...U Choose the upper item on the menu list 19 MENU Show Hide the menu 20 MENU Choose the lower item on the menu list 21 OPTION the key for DSO setting and combines with four arrow keys Then to set main time base trigger horizontal position and trigger vertical position Furthermore the keys possible to adjust the display multiplying factors of M waveform CHMath Volts Div and display vertical position...

Page 18: ...ower via AC DC adapter as shown in Figure 1 The oscilloscope may still work with built in Li ion battery even without AC power supply Turn the oscilloscope on by pressing down the power on off key The instrument then performs self checking after power on A greeting window and a sentence Press any key to continue will display on the screen when the system finishes self checking The users can press ...

Page 19: ...he following information Auto The oscilloscope is working in the automatic mode and displaying the waveform under the non trigger state Trig d The oscilloscope has detected a trigger and collecting the information generated after the trigger Ready All pre triggered data have been captured and the oscilloscope has been ready to receive trigger signals Scan The oscilloscope can gather and display th...

Page 20: ... display area Red waveform represent CH1 blue waveform represent CH2 20 The red pointer gives the grounding datum point of the waveform on CH1 which is the zero position of CH1 No display of this pointer indicates that the channel has not been opened 6 4 Menu Description The following example shows how to use the tool s menus to select a function as shown in the following figure 1 Press the MENU k...

Page 21: ...Zero is displayed blue and blue can be used to adjust the vertical position of CH2 the relevant information will be displayed 3 When CH2 Volts Div is displayed blue and blue can be used to adjust the voltage scale of CH2 the relevant information will be displayed 6 5 2 Setting the Horizontal System and Trigger Position OPTION key is a multiple setting key to set trigger vertical position main time...

Page 22: ... yellow to adjust the main time base press yellow or yellow to adjust trigger level position press R key to zero the trigger level position 3 Press OPTION again and left bottom side display as Time Trig See the following figure 8 Figure 8 4 Press yellow or yellow to adjust time base horizontal position press V key to zero the 16 ...

Page 23: ...lternative triggering 1 Press OPTION and left bottom display as Time Trig 2 or Trig 1 See the following figure 9 Figure 9 Adjust Time and Trig 2 2 Press yellow or yellow to adjust time base horizontal position press V key to zero the horizontal position press yellow or yellow to adjust trigger level position in Channel 2 or Channel 1 press R key to zero the trigger level position 3 Press OPTION ag...

Page 24: ...tment of which you can regulate the display position of the waveform on the screen z Main time base It means the time values represented by a division in the horizontal direction of the display area z Trigger horizontal position It means the time deviation between the actual trigger point and the screen central line which will be displayed as 0 at the center point of the screen z Trigger level pos...

Page 25: ... and CH2 for scope measurements four safety 4 mm banana jack inputs for Multimeter R V and A measurements Isolated input allows independent floating measurements between Multimeters and Scopes 6 8 Displaying an Unknown Signal with Auto Set The Auto Set feature lets the Oscilloscope display and measure unknown signals automatically This function optimizes the position range time base and triggering...

Page 26: ...readings are selectable independently and the measurements can be done on the input CH1 or input CH2 waveform To choose a frequency for CHI1 do the following 1 Press MENU key and the function menu appears on the right side of the screen 2 Press MENU or MENU key to select MEAS SET1 Five items selectable are visible at the bottom of the screen 3 Press F1 key and select Freq CH1 from the mean square ...

Page 27: ...ts 6 11 Freezing the Screen You can freeze the screen all readings and waveforms 1 Press the RUN STOP key to freeze the screen and STOP appears at top right side of the screen 2 Press the RUN STOP key once more to resume your measurement See the following figure 13 Figure 13 Freezing the Screen 21 ...

Page 28: ... items selectable displayed at the bottom of the screen 3 Press the F3 key to select Average then press F4 key to jump to Averages 16 item This averages the outcomes of 16 acquisitions and shows the final averaging result on the screen shown as the following figures See the following figure 14 Figure 14 Average Factor Sampling Mode 6 13 Using Persistence to Display Waveforms You can use Persistenc...

Page 29: ...function to display events glitches or other asynchronous waveforms of 50 ns or wider 1 Press MENU key and the function menu appear at the right side of the screen 2 Press MENU or MENU key to select the ACQU MODE Four items selectable are displayed at the bottom of the screen 3 Press F2 key and jump to Peak Detect In this case you can test the glitch Now you can see a screen that looks like the fo...

Page 30: ...ys the averaged waveform in this mode by which the random noise can be reduced Duration time When a new waveform is displayed the previous waveform shown on the screen does not disappear immediately only to be displayed for a period of time that is the duration time by setting which the waveform can be displayed more continuously and thus a display similar to that shown by the analog oscilloscope ...

Page 31: ... Press the MENU key and the function menu appears at the right side of the screen 2 Press the MENU or MENU key to select CH1 SETUP Four items selectable are displayed at the bottom of the screen 3 Press F4 key to jump to Inverted set it as ON The inverted waveform of CH1 is displayed on the screen Now you can see a screen that looks like the following figure 18 Figure 18 Inverted On 25 ...

Page 32: ...do the following 1 Press the MENU key and the function menu is displayed at the right side of the screen 2 Press the MENU or MENU key to select the WAVE MATH Five items selectable appears at the bottom of the screen 3 Press F3 key to select CH1 CH2 and the calculated waveform M green appears on the screen 4 Press OPTION until left bottom display as CHMath Volts Div CHM Zero 5 Press yellow or yello...

Page 33: ...e is vector format and another is bitmap format according to the display settings from the choice of the communication settings Then the file name in sequence will be WAVE1 BIN WAVE2 BIN WAVE3 BIN or WAVE1 BMP WAVE2 BMP WAVE3 BMP Then connect the USB mass storage device with the computer after data is saved and open the data of vector format with OWON analysis software or direct open the bitmap No...

Page 34: ...us and perform basic operations 7 2 Making Meter Connections Use the four 4 mm safety banana jack inputs for the Meter functions 10A mA COM V Ω C See figure 2 on P8 for the connections 7 3 Multimeter Operation Window Figure 20 Multimeter Operation Window Description 1 Battery electric quantity indicator 2 Manual Auto range indicators among which the MANUAL means measuring range in manual operation...

Page 35: ...To multiply the reading of dial pointer by multiplying power will get the measurement result 8 The main reading of measurement 9 Automatic control measuring range 10 Absolute relative magnitude measuring control The sign expresses the absolute magnitude measuring control and represents the relative magnitude measuring control Manually measuring range control 11 Manual measurement control 12 Test l...

Page 36: ...wing figure 21 Figure 21 Resistance Measurement 7 4 2 Measuring Diode To make a measurement on the diode do the following 1 Press the R key and R appears at the top of the screen 2 Press SET key until the following is displayed on the screen 3 Insert the black lead into the COM banana jack input and the red lead into the V Ω C banana jack input 4 Connect the red and black leads to the resistor and...

Page 37: ...is shown on the screen 3 Insert the black lead into the COM banana jack input and the red lead into the V Ω C banana jack input 4 Connect the red and black leads to the test point If the resistance value of the tested point is less than 50Ω you will hear beep sound from the test tool Now you can see a screen that looks like the following figure 23 Figure 23 On off Test 7 4 4 Measuring Capacitance ...

Page 38: ...e relative value measuring mode to improve measuring precision It will take about 30seconds if capacitance measurement is larger than 40uF Now you can see a screen that looks like the following figure 24 Figure 24 Capacitance Measurement 7 4 5 Measuring DC Voltage To measure a DC voltage do the following 1 Press the V key and DCV appears at the top of the screen 2 Insert the black lead into the CO...

Page 39: ...en 2 Press the SET key and ACV appears at the top of the screen 3 Insert the black lead into the COM banana jack input and the red lead into the V Ω C banana jack input 4 Connect the red and black leads to the measured points and the AC voltage values of measured points will be displayed on the screen Look at the display you can see a screen that looks like the following figure 26 Figure 26 AC Vol...

Page 40: ...easured points will be displayed on the screen Look at the display you can see a screen that looks like the following figure 27 Figure 27 DC Current Measurement for 400 mA To measure a DC current which is larger than 400 mA do the following 1 Press the A key and DCA appears at the top of the screen The unit on the main reading screen is mA 2 Press F5 key change to 10A measurement the unit on the m...

Page 41: ...bottom of screen press F4 and F5 to switch the measurement between mA and 10A 400mA is acquiescently 2 Press the SET key once and ACA is visible at the top of the screen 3 Insert the black lead into the COM banana jack input and the red lead into the mA banana jack input 4 Connect the red and black leads to the measured point and the AC current value of the measured point will be displayed on the ...

Page 42: ...ert the black lead into the COM banana jack input and the red lead into the 10A banana jack input 5 Connect the red and black leads to the measured point and the AC current value of the measured point will be displayed on the screen 6 Press F4 return to 400mA measure Look at the display you can see a screen that looks like the following figure 30 Figure 30 AC Current Measurement for 10A 7 5 Freezi...

Page 43: ...cquire a reference value 1 Press R key and R is displayed on the top side of the screen 2 Press the SET key until C appears at the top of the screen 3 Plug capacitance extended module in capacitance measure jack 4 When the reading leveling off press F2 key and is displayed on the top side of the screen The saved reference value is displayed below 5 Plug capacitor the displayed major reading on the...

Page 44: ...uring range is increased by a stage when pressing F1 key each time and when reaching the highest stage it jumps to the lowest stage by pressing F1 key once again To multiply the reading of dial pointer by multiplying power and the unit of main reading on the screen will get the measurement result 3 Press F3 key and AUTO is displayed on the top left side of the screen to switch back to the automati...

Page 45: ...nt vertical menu and each item can be set respectively based on the specific channel To make vertical CH1 and CH2 settings do the following 1 Press the MENU key and the function menu appears at the right of the screen 2 Press the MENU or MENU key to jump to CH1 SETUP and 4 options appears at the bottom of the screen 3 Select key from F1 to F4 to make different settings Now you can find a screen th...

Page 46: ...re a correct vertical scale reading Probe OFF ON Waveform is displayed normally Open the Invert function of the waveform setting Inverted 8 2 1 Setting the Channel Coupling Take CH1 for example Press F1 key set Coupling as AC to make an AC coupling setting The DC component contained in the tested signal is blocked Press F1 key set Coupling as DC to make a DC coupling setting Both DC and AC compone...

Page 47: ...o make a Close setting on CH1 Press F2 Channel key first and then ON to make an Open setting on CH1 8 2 3 Setting the probe attenuation To prevent excessive input voltage we recommend you to set the probe attenuation level to the 10X position to prevent excessive voltage After attenuating the probe level by 10 1 you also need to magnify the display level by 10 X to match the displayed amplitude wi...

Page 48: ...veform can be adjusted with CHM VOL which is displayed in the scale factor form The amplitude ranges from 0 001 through 10 and steps in the 1 2 5 form that is it can be expressed as 0 001X 0 002X 0 005X 10X The position of the calculated waveform can be adjusted up and down with the CHM ZERO key used The corresponding operation function table Setting Description CH1 CH2 CH1 waveform minus CH2 wave...

Page 49: ...arting the acquisition of data the oscilloscope collects sufficient data to draw the waveform at the left side of the triggering point With waiting for the triggering condition the oscilloscope is gathering data continuously After a trigger is detected the oscilloscope gathers enough data continuously to draw the waveform at the right side of the triggering point To make a trigger mode setting do ...

Page 50: ...oming signal edge Use the edge trigger for all signals except for video related ones Video trigger Perform video field trigger or line trigger on the standard video signals Alternate trigger When the signal frequency different in Channel 1 and 2 it can also guarantee the stability approaching in two channels The following describes Edge trigger Video trigger and Alternating trigger menus respectiv...

Page 51: ...ed and then stop sampling Trig mode 1 2 To next menu AC DC HF Rjc LF Rjc With this mode selected the DC component is prevented from passing though All components are allowed The HF part of the signal is prohibited and only the LF component is allowed The LF part of the signal is prohibited and only the HF component is allowed Coupling To go to holdoff menu Holdoff 2 2 Back to previous menu 8 5 2 V...

Page 52: ...a video odd field trigger synchronization setting Make a video even field trigger synchronization setting Make a video designed line synchronization setting Sync 1 2 To next menu The Video triggering menu Second page 1 When the sync is Line Field Odd Field Even Field the second page menu is shown as bellow NTSC PAL SECAM Video format setting MODU Holdoff To go to holdoff menu 2 2 Back to previous ...

Page 53: ... the type is set as Edge trigger Settings Description Function menu CH1 CH2 Setting trigger type and others info for Channel 1 Setting trigger type and others info for Channel 2 CH SEL Edge Video Set vertical channel trigger as edge trigger Set vertical channel trigger as video trigger Type Rising Falling Triggering on the rise edge of the signal Triggering on the fall edge of the signal Edge type...

Page 54: ...e Set synchronous trigger in video even line Set synchronous trigger in video Line NUM when the sync is Line Field Odd Field Even Field the following menu is shown as below MODU Modulation NTSC PAL SECAM Set synchronization and account choose video standard Set synchronization and account choose video standard Holdoff To go to holdoff menu when the sync is Line NUM the following menu is shown as b...

Page 55: ...d in this mode in which it will be triggered compulsively when waiting for a specified period of time without any triggering condition ignited When an invalid trigger is enforced the oscilloscope can not keep the waveform in phase Normal trigger mode In this mode the oscilloscope cannot acquire the waveform till it is triggered When there is not any trigger the oscilloscope will display the origin...

Page 56: ...e The persistence setting sets how long the old waveforms remain in the display useful for observing the waveform variations Persist YT XY Display the relative relationship between vertical voltage and horizontal time Display CH1 on the horizontal axis and CH2 on the vertical axis Format Bitmap Vectors The data transmitted in communication are bitmaps The data transmitted in communication are vect...

Page 57: ...as X axis and CH2 input as Y axis This display mode is convenient for viewing the phase relationship between CH1 and CH2 when the oscilloscope is under the sampling mode in which no trigger is found the data appear in light spots Operations for various control keys are shown as below The CH1 Volts Div and CH1 Zero for CH1 see System on P15 are used to set the horizontal scale and position The CH2 ...

Page 58: ...of two channels To make a cymometer setting do the following If you want to measure frequency of the two Channels you can do as the follows 1 Press MENU the function menu will appear on the right of the screen 2 Press MENU or MENU and choose DISP SET five options will show at the bottom of the screen 3 Press F5 and choose ON See the following figure 48 Figure 48 Cymometer 8 8 Waveform Saving Setup...

Page 59: ...following 1 Press the MENU key and the function menu appears at the right of the screen 2 Press the MENU or MENU key to select the WAVE SAVE Four options are displayed at the bottom of the screen 3 Press the F1 key to select the signal source CH1 4 Press the F2 key to select the address A 5 Press the F3 key to save the waveform on CH1 in address A To display the saved waveform on the screen do the...

Page 60: ... save a waveform on CH1 in address A do the following 1 Press the MENU key and the function menu appears at the right of the screen 2 Press the MENU or MENU key to select the WAVE SAVE Four options are displayed at the bottom of the screen 3 Press the F1 key to select FFT 4 Press the F2 key to select the address A 5 Press the F3 key to save the waveform on CH1 in address A To display the saved wav...

Page 61: ...on the display asking you to remove all cables and probes from oscilloscope 3 After removing all cables press F2 Auto calibration again The Auto calibration automatically starts and a message appears showing that the calibration is ongoing To interrupt calibration press any key during the calibration 8 10 Making Automatic Measurements The oscilloscope can perform 20 types automatic measurements su...

Page 62: ...e the Rise Time of CH2 Rise Time CH1 CH2 Measure the Fall Time of CH1 Measure the Fall Time of CH2 Fall Time CH1 CH2 Measure the Width of CH1 Measure the Width of CH2 Width CH1 CH2 Measure the Width of CH1 Measure the Width of CH2 Width CH1 CH2 Measure the Duty of CH1 Measure the Duty of CH2 Duty CH1 CH2 Measure the Duty of CH1 Measure the Duty of CH2 Duty CH1 CH2 Measure the DelayA B of CH1 Measu...

Page 63: ...ou to make manual cursor measurements on time and voltage The signal sources include Channel 1 CH1 Channel 2 CH2 8 11 1 Setting the cursor measurement in normal mode The cursor measurement menus are listed and described in the following table Function menus Settings Description OFF Voltage Time Close the cursor measurement Display the voltage measurement cursor and menu Display the time measuremen...

Page 64: ... Press the F2 key to select the measured channel CH1 5 Press OPTION and display as Cursor 1 Cursor 2 When setting OPTION or OPTION V2 will move up and down and the relating voltage value to the zero position of Channel 1 will show at the bottom of the screen Setting OPTION or OPTION V1 will move up and down and the relating voltage value to the zero position of Channel 1 will show at the bottom of...

Page 65: ...ment type as Time Two vertical dashed lines T1 and T2 appear on the screen 4 Press the F2 key and jump to the measured channel CH1 5 Press OPTION and display as Cursor 1 Cursor 2 Press yellow or yellow and see T1 move left and right and LCD display the time value match T1 to point position in panel press yellow or yellow and see T2 move left and right and LCD display the time value match T2 to poi...

Page 66: ...e Source CH1 CH2 Display the selected source in the FFT MODE menu unable to change here Display the difference of measured value between the two channels Delta Curs1 Display the relating measured value of Cursor 1 Curs2 Display the relating measured value of Cursor 2 To make a range measurement on CH1 doing the following 1 Press the MENU key and the function menus are displayed at the right of the...

Page 67: ...Figure 56 Cursor range measurement 8 12 Autoscale The function is applied to follow up signals automatically even if the signals change at any time Autoscale enables the instrument to set up trigger mode voltage division and time scale automatically according to the type amplitude and frequency of the signals The menu is as follows Function menus Settings Description OFF ON Turn off Autoscale Turn...

Page 68: ...e ware when turning on Autoscale 1 The symbol is flickering on the top left corner of screen every half second 2 The oscilloscope is setting Triggers automatically for the incoming signal based on its best approach Access from front panel to trigger controls is disabled A warning message will be displayed on the bottom left corner of the screen when trying to change submenus of Trigger controls 3 ...

Page 69: ...odes Record Play back Storage OFF Record Record the waveforms up till the set number of frames at specified time intervals Record menu list Menu Setting Instruction Record Play back Storage OFF Select record mode Select play back mode Select storage mode Turn off all recorder functions Mode End frame 1 1000 Set number of record frames Play Stop Press to start recording Press to stop recording Oper...

Page 70: ...ame cur frame and end frame The value decreases at the mode of setting start frame cur frame and end frame Direction 2 2 Back to previous menu Note the Run Stop button can also replay or continue the waveform display Storage Store recorded waveforms in non volatile memory according to the setup frames Storage menu list Menu Setting Instruction Start Frame 1 1000 Set first frame to be saved End fra...

Page 71: ... 3 Press the F1 key to select the Record mode 4 Press F2 and set the End frame at 300 select the direction as Increase or Decrease in the second menu before step4 5 Press F4 and set the time Interval to 1 00ms 6 Press F5 to enter into the second menu 7 Press F1 and select OFF to stop refreshing waveforms 8 Press F5 to back up to the first menu 9 Press F3 and select Play to operate Shown as Fig 58 ...

Page 72: ...breaks down sign The following table describes the FFT menu Function Menu Setting Instructio ON OFF Turn on FFT function Turn off FFT function FFT Source CH1 CH2 Select CH1 as FFT source Select CH2 as FFT source Window Rectangle Blackman Hanning Hamming Select window for FFT Format dB Vrms Set Vrms as vertical scale unit Set dBVrms as vertical scale unit Zoom 1 2 5 10 Set multiple 1 Set multiple 2...

Page 73: ...3 to switch over Window including Hamming Rectangle Blackman and Hanning 6 Press F4 to swit 7 Press F5 to zoom in or out the waveform the magnification includes 1 2 5 10 8 If the FFT source is CH1 press VOLTS PO z One of the following three prompts circularly will show at the bottom left of the screen when the format is dB CH1 Volts Div FFT Zero FFT dB Scale One of the following when the format is...

Page 74: ... displays FFT Zero press red and red button t as FFT 1 20 divs which means the pink cursor departures from the center line for 1 20 DIV press VOLTS POSITION red to display FFT dB Scale CH1 20dB shows on the left bottom that is the arithmetic product is 24 0dB The operation steps are the same as FFT source CH2 10 Press OPTION yellow the following prompts shows on the left bottom on the scree Time B...

Page 75: ...measuring the frequency spectrum of nonrepetitive signals and measuring frequency components near DC Use rectangle for measuring transients or bursts where the signal level before and after the event are nearly equal Also use this window for equal amplitude sine waves with frequencies that are very close and for broadband random noise with a relatively slow varying spectrum Rectangle Hamming This ...

Page 76: ...transients or bursts where the signal levels before and after the event are significantly different Hanning This is the best window for measuring the amplitude of frequencies but worst at resolving frequencies Use Blackman Harris for measuring predominantly single frequency waveforms to look for higher order harmonics Blackman Fig 61 62 63 64 show four kinds of window function referring to sine wa...

Page 77: ...hat have a DC component or offset can cause incorrect FFT waveform component magnitude values To minimize the DC component choose AC Coupling on the source signal To reduce random noise and aliased components in repetitive or single shot events set the oscilloscope acquisition mode to average Term interpretation Nyquist frequency The highest frequency that any Real Time Digital Oscilloscope can me...

Page 78: ...ord 7 Measure 1 and Measure 2 8 15 System State Menu The system state menu is used to display information about the present horizontal system vertical system trigger system and others The operation steps are shown as below 1 Press the MENU key and the function menu is displayed at the right of the screen 2 Press the MENU or MENU key to select the SYS STAT Four options appear at the bottom of the s...

Page 79: ...ect be deepened the value can be adjusted 5 Press OPTION or OPTION to change the corresponding value of the year month date hour minute and second effectively 6 Press OPTION or OPTION to switch to the year month date hour minute and second that are in need of adjustment 7 Press OPTION to enter into normal time mode after timing 8 If need to timing please kindly repeat the above steps Note The digi...

Page 80: ...ellow and yellow key to adjust the time base window area defined by two cursors the window size will vary 5 Press OPTION key and call TIME at this time press yellow and yellow to adjust the window position defined by two cursors the window position is the time difference of the window center to main time base s horizontal pointer 6 Press F3 key select Window the defined window extends into the ful...

Page 81: ...ss MENU key display the function menu on the right side of the screen 2 Press MENU or MENU key to select DISP SET 3 Press F4 key select the Bitmap or Vectors for data transmission 4 Use a data line to connect the oscilloscope and the PC 5 Open the OWON software that had been installed completely 6 Setting the parameter under the use s manual and then starts data transmission ...

Page 82: ...ler than the real value under the oscilloscope mode Check whether the channel attenuation level matches the actual probe attenuation ratio 5 The waveform is displayed on the screen but is not stable under the oscilloscope mode z Check whether the signal item in the trigger mode menu matches the actually used signal channel z Check on the trigger mode The edge trigger mode is applicable to the univ...

Page 83: ... the system function menu must be opened to make the system perform a Auto calibration procedure see calibration on P55 Except those specifications marked with the word Typical all specifications can be up to Sampling Normal sampling Peak detection Average value Sampling modes 100 MS s HDS1022M N 500 MS s HDS2062M N Max 1 GS s HDS3102M N Sampling rate Input Input coupling DC AC Ground 1MΩ 2 connec...

Page 84: ...0mV div 40V 200mV div 5 V div HDS2062M N HDS3102M N Displacement range 20 MHz HDS1022M N 60 MHz HDS2062M N 100MHz HDS3102M N Analog bandwidth Single bandwidth Full bandwidth Low frequency response AC coupling 3dB 5Hz at the BNC 17 5ns HDS1022M N 5 8 ns HDS2062M N 3 5 ns HDS3102M N Rise time typical one at the BNC DC gain accuracy 3 The voltage difference V between any two points on the waveform af...

Page 85: ...ng the cymometer will not work Single source Measurement Cursor measurement Voltage difference V and time difference T between cursors Peak to peak value average value root mean square value frequency period Vmax Vmin Vtop Vbase Vamp Overshoot Preshoot Rise Time Fall Time Width Width Duty Duty DelayA B and DelayA B Auto measurement Probe 1X position 10X position Bandwidth Up to 6 MHz DC Up to full...

Page 86: ...irect Current DC Range Accuracy Resolution 40 00mA 1 5 1 digit 10uA 400 0mA 1 5 1 digit 100uA 10A 3 3 digits 10mA Alternating Current AC Range Accuracy Resolution 40 00mA 1 5 3 digits 10uA 400 0mA 2 1 digit 100uA 10A 5 3digits 10mA Resistance Range Accuracy Resolution 400 0 Ω 1 3 digits 0 1Ω 4 000KΩ 1Ω 40 00KΩ 10Ω 400 0KΩ 100Ω 1 1 digit 4 000MΩ 1KΩ 40 00MΩ 1 5 3 digits 10KΩ Capacitance Range Accur...

Page 87: ...zontal 480 vertical pixels Display color 65536 colors Power Adapter Power supply 100 240 V AC 50 60Hz Power output 1 5VDC HDS1022M N 9 VDC HDS2062M N HDS3102M N Current output 1500mA HDS1022M N 3000mA HDS2062M N HDS3102M N Battery type 7 4V built in Li ion battery Working environment Temperature Used battery 0 to 50 C 32 to 122 F Operation Power adapter 0 to 40 C 32 to 104 F Storage 20 to 60 C 4 t...

Page 88: ... battery before storage Charging the oscilloscope The lithium battery is possibly not charged when delivery To make the battery with enough electric quantity it must be charged for 4 hours the test tool must be turned off during charging The battery can supply power for 4 hours after being charged completely When supplying power by using the battery a battery indicator is displayed on the top of t...

Reviews: