ERROR MESSAGES
8.1. ERROR MESSAGES
An anomaly is indicated as follows:
- emission of an intermittent acoustic signal (beep)
- MALFUNCTIONING INDICATOR LED intermittently turns on
- the error code (E ….) appears on the display)
- all control panel functions are inhibited
In this case turn off the timer and then turn it back on. If the error persists, refer to the following table.
The following chart gives a list of error messages that may appear while
Owandy-RX DC
radiographic system is working.
The chart also includes the causes of the error messages to be communicated to the technical support and, in some
cases, what to do to solve them.
If the error is not fi xed by carrying out the described solution, turn off the power supply from the electrical panel (wall
mounting version) or unplug the system (mobile version)
ERROR
MESSAGES
CAUSE
SOLUTION
E00
RX1 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
E01
RX2 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
E02
CORRUPTED EEPROM DATA
CALL THE “ASSISTANCE SERVICE”
E03
EEPROM DATA NOT SAVED PROPERLY
CALL THE “ASSISTANCE SERVICE”
E07
LINE VOLTAGE VALUE NOT INCLUDED
WITHIN THE ±10% NOMINAL VALUE
CALL THE “ASSISTANCE SERVICE”
E08
THE X-ray KEY
ALWAYS SEEMS TO BE PRESSED
MAKE SURE IT IS NOT JAMMED
E09
ANOMALY IN THE CONTROL PANEL
CALL THE “ASSISTANCE SERVICE”
E12
THE EXPOSURE
HAS BEEN PREMATURELY INTERRUPTED
KEEP THE X-ray KEY
PRESSED TILL THE END OF THE
EXPOSURE
E20
ANOMALY IN THE TRIAC/RELAY
CALL THE “ASSISTANCE SERVICE”
E21
ANOMALY IN THE ELECTRONIC CIRCUIT
CALL THE “ASSISTANCE SERVICE”
E23
INCORRECT DIP-SWITCH CONFIGURATION
CALL THE “ASSISTANCE SERVICE”
E24
THE CONTROL BUTTON DOES NOT
CORRESPOND
TO THE SELECTED TUBEHEAD
CALL THE “ASSISTANCE SERVICE”
E30
THE TUBEHEAD
DOES NOT WORK PROPERLY
CALL THE “ASSISTANCE SERVICE”
E32
THE TUBEHEAD
IS NOT IN THE CORRECT MODE
CALL THE “ASSISTANCE SERVICE”
E33
THE TUBEHEAD
HAS NOT COMPLETED THE EXPOSURE
REPEAT THE EXPOSURE WITHOUT THE
PATIENT
CALL THE “ASSISTANCE SERVICE”
E40
PROBLEM IN THE FREQUENCY
OR REGULATION
CALL THE “ASSISTANCE SERVICE”