OPERATOR’S MANUAL • Owandy-RX DC • 02/2020 • NXDCEN160A
Operator’s manual – Owandy-RX DC
PLEASE NOTE
60kV mode can be selected with the
Owandy-RX DC
radiographic system only.
PLEASE NOTE
After the modifi cation, default exposure values will be automatically changed.
STEP 5 : check the selected radiographic current
The LED of the radiographic current should be turned on
LED
4mA ON
indicates that the radiographic system is set with reduced dose.
It is advisable the use of digital radiographic technique.
LED
8mA ON
indicates that the radiographic system is set with nominal dose.
It is advisable the use of conventional radiographic technique.
to change the selection press again the button
PLEASE NOTE
The 4mA mode can be selected with the
Owandy-RX DC
radiographic system only.
PLEASE NOTE
After the modifi cation, default exposure values will be automatically changed.
STEP 6 : check the selected type of patient
The LED of the desired type of patient should be turned on
LED
CHILD ON
indicates that the radiographic system is set for a patient with a small physique
LED
ADULT ON
indicates that the radiographic system is set for a patient with a large physique
to change the selection press again the button