16. LOOP TEST (SIMULATIVE OUTPUT) FUNCTION
With this function, simulative output of analog, pulse and status is carried out for verifying the connection
between this flowmeter and external instruments. The loop test is operated from the "LP.tSt" submenu under
the Setup mode.
■
Analog simulative output
Outputs a desired current (2.4mA to 21.6mA) from the analog output terminal.
■
Pulse simulative output
Outputs a pulse of desired frequency from the pulse output terminal.
■
Status simulative output
Outputs ON or OFF from the status output terminal simulatively.
16.1 Simulative Output Operation
Select the item to be output simulatively from the "LP.tSt" submenu of the Setup menu.
Both pulse and status simultative output operations can be performed by the same procedure.
When 10 minutes pass without any key operation during output operation, output stops and the Measure mode
is restored.
Simulative output (Analog output)
SEL
: Changes numeral
Press and hold SEL : Cancel
ENT
: Shifts digit
Press and hold ENT : Fix
Confirm the entry by "Press and hold ENT".
After Confirmation, "Out" blinks and the set
current is output.
"Press and hold SEL" stops output and returns to the original screen.
Set the current value of output.
Press and
hold ENT
Press and
hold ENT
Press and
hold SEL