12.5 Display Transition (5) Setting Diagnosis and Maintenance Functions
Parameter third item
Content of parameter
Default
Setting range
On
On
On
Off
Monitors the current flowing through the coil exciting circuit and diagnoses for abnormality.
* This setting is valid in the auto self-diagnosis in normal measurement.
When diagnosis is started from Menu or through communication, diagnosis is carried out
regardless of this setting.
Diagnoses the circuit voltage in the transmitter.
* This setting is valid in the auto self-diagnosis in normal measurement.
When diagnosis is started from Menu or through communication, diagnosis is carried out
regardless of this setting.
Diagnoses whether or not noise, etc. is superimposed on flow measurement signal.
* This setting is valid in the auto self-diagnosis in normal measurement.
When diagnosis is started from Menu or through communication, diagnosis is carried out
regardless of this setting.
This is set for identifying the instruments when two or more instruments are connected in
multiple points. Set the parameter so that instruments with duplicated address are not present.
In the case of 1 to 1 connection, leave the parameter at "0".
Diagnoses whether or not electrode of probe is wetted and diagnoses the state of error
entry in the pipe under measurement.
* This setting is valid in the auto self-diagnosis in normal measurement.
When diagnosis is started from Menu or through communication, diagnosis is carried out
regardless of this setting.
On, OFF
On, OFF
On, OFF
On, OFF
Content of parameter
Default
Setting range
Content of parameter
Default
Setting range
―
19200
0
0 to 63
(2 integer digits)
Not disclosed
"SEL"
"Press and hold SEL"
"Press and hold ENT"
Other
ENT
Displays instant flow rate
Total flow
Displays % flow rate
Resettable total flow
Displays flow velocity
Setup mode
Circuit voltage monitoring
Coil exciting circuit
error diagnosis
CoiL.S
Set flow parameter 2
F.Prm2
Set flow parameter 3
F.Prm3
Parameter major item
[Top row display] Selects display
[Bottom row display] Displays total flow
Enter password
00-00
01-00
02-00
03-00
04-00
05-00
06-00
07-00
08-00
09-00
10-00
11-00
12-00
13-00
14-00
PAS.wd
11-01
11-02
Set display
dSP.St
Set flow parameter 1
F.Prm1
Set self-diagnosis
diAG.S
Loop test
LP.tSt
Analog output trim function
A.trim
Self-check function
CHECK
Set analog output
A.Out
Set pulse output
P.Out
Set error output
Er.Out
Set alarm
ALArm
12-01
12-02
12-03
Pulse fixed output
LP.PLS
Status fixed output
LP.StS
Analog fixed output
LP.AnA
Set status output
S.Out
Select status output object
St.SEL
Parameter secondary item
Cirt.S
Measurement error
signal diagnosis
11-03
SiGL.S
Empty pipe detection
diagnosis
11-04
Enpy.S
RS485 communication
speed
Polling address
PoL.Ad
14-03
14-04
bAu.br
LCD check
14-05
ALL.on
Maintenance and
calibration mode
14-06
otHEr
Circuit voltage diagnosis
Coil, exciting circuit
diagnosis
Coil.C
14-02-01
14-02-02
Cirt.C
Measurement signal
diagnosis
14-02-03
SiGL.C
Empty detection function
14-02-04
EmPy.C
Analog output 4mA
output adjustment
13-01
Analog output 20mA
output adjustment
13-02
trm.20
trm.04
Self-diagnosis, overall
14-01
Self-diagnosis, individual
14-02
mAnu
Auto
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
―
Content of parameter
Default
Setting range
Outputs the specified current from analog output.
Outputs pulse of the specified frequency from pulse output.
Outputs the specified level from status output.
Adjusts analog 4mA output.
Adjusts analog 20mA output.
Diagnoses the circuit voltage in the transmitter.
Monitors the current flowing in coil exciting circuit and diagnoses for any
abnormality.
Diagnoses whether or not noise, etc. are superimposed on flow measure-
ment signal.
Diagnoses whether or not the electrode of probe is wetted and judges the
error entry state in the pipe under measurement.
Sets the baud rate of RS-485 communication for setting.
Lights all segments of LCD.
Mode dedicated to OVAL service personnel
Content of parameter
Default
Setting range
Executes all self-diagnosis items provided and displays the results.
Content of parameter
Default
Setting range
9600bps
,
19200bps
,
38400bps
m/s
modE
**.***
*****
*****
**.**
**.**
Parameter name
Layer code (Major item ― Secondary item ― Third item)
Parameter
display symbol