12.3 Display Transition (3) Setting Output Signal
4mA output setting
20mA output setting
Analog output damping
20mA
F.Prm2
F.Prm3
00-00
01-00
02-00
03-00
04-00
05-00
06-00
07-00
08-00
09-00
10-00
11-00
12-00
13-00
14-00
PAS.wd
05-01
05-02
05-03
dSP.St
F.Prm1
diAG.S
LP.tSt
A.trim
CHECK
A.Out
P.Out
Er.Out
ALArm
06-01
06-02
06-03
06-04
06-05
Frequency factor
Fq.FAC
Flow rate factor
rt.FAC
Pulse output low cut
Lo.Cut
Pulse weight
wEit
Pulse calculation mode
P.modE
S.Out
St.SEL
Parameter secondary item
A.dAmP
Analog output low cut
05-04
Lo.Cut
4mA
Status output function
07-01
Status output mode
07-02
S.modE
S.FunC
Content of parameter
Default
Setting range
0 to Measurement range setting
* Flow (5 significant 1 exponent digit)
0.0 to 100.0s (seconds)
0.0
0.0
As
specified
As
specified
As
specified
As
specified
As
specified
0.0000E0
0 to Measurement range setting
* Flow (5 significant 1 exponent digit)
Can change the analog output lower limit flow (flow at 4mA output).
When you change the measurement range, change this parameter as necessary.
* Unit is subject to the setting of instant flow rate display unit.
Can change the analog output upper limit flow (flow at 20mA output).
When you change the measurement range, change this parameter as necessary.
* Unit is subject to the setting of instant flow rate display unit.
Can change the time constant for analog output damping.
This affects the flow measurement damping "F.dAmP" in double.
* Time constant refers to the time instruction reaches approx. 63% in step response.
Fixes the output to 4mA until a certain amount of flow is reached.
It is set in % to max. measurement flow.
* For details, refer to "ANALOG OUTPUT FUNCTION".
0.0 to 10.0%
Content of parameter
Default
Setting range
On ACt
no.FunC
Sets the allocation of status output.
* For details, refer to "15. STATUS OUTPUT FUNCTION".
Sets status output logic.
* For details, refer to "15. STATUS OUTPUT FUNCTION".
no.FunC : Status output not used
Err.StS : Error, alarm, status output
HL.ALA : Higher and lower limit alarm output
FLo.dir : Flow direction
OFF.ACt : Off, Active
On.ACt : On, Active
Content of parameter
Default
Setting range
0.0
Can select pulse output mode. (Frequency pulse or Factored pulse)
* Note that changing this parameter resets the total value.
Sets the full-scale frequency of "Frequency pulse".
* This is not displayed when "Factored pulse" is selected.
Sets the flow rate at the full-scale frequency output of "frequency pulse".
* This is not displayed when "Factored pulse" is selected.
Stops output pulse until a certain amount of flow is reached.
Sets by the rate of pulse output to full-scale flow.
* This is not displayed when "Factored pulse" is selected.
Sets the flow rate per one pulse of "Factored pulse".
Unit is subject to the setting of total flow unit.
* This is not displayed when "Frequency pulse" is selected.
"FrEq" (Frequency pulse)
"WEit" (Factored pulse)
1 to 10000 Hz
0 to Measurement range set flow
* 5 significant exponent
0.0 to 10.0%
Set the flow rate so that the frequency
at max. flow rate is within 10,000 Hz.
* (5 significant exponent)
(Note)
(Note)
(Note)
(Note)
m/s
modE
%
**.***
**.**
**.**
NOTE: Changing these parameters resets the total value.
"SEL"
"Press and hold SEL"
"Press and hold ENT"
Other
ENT
Displays instant flow rate
Total flow
Displays % flow rate
Resettable total flow
Displays flow velocity
Setup mode
Set flow parameter 2
Set flow parameter 3
Parameter major item
[Top row display] Selects display
[Bottom row display] Displays total flow
Enter password
Parameter name
Layer code (Major item ― Secondary item ― Third item)
Parameter
display symbol
Set display
Set flow parameter 1
Set self-diagnosis
Loop test
Analog output trim function
Self-check function
Set analog output
Set pulse output
Set error output
Set alarm
Set status output
Select status output object
*****
*****