L-770HP--E
E-454-1-E
5
3. Installation of converter (PA5K) / procedure of wiring
3.1
Procedure of installation
(1)
Select a place where there is few mechanical vibration and corrosive gas.
(2)
Select a place where there the humidity is low and the temperature is almost ordinary temperature.
(3)
To install more than one converter under an environment where the ambient temperature of converter
exceeds 40°C, the attachment pitch should be 60mm or more.
(4)
Install the product at a place where you can carry out the maintenance easily.
(5)
Attach the converter with the attachment dimensions
Fig.3.1 Converter external dimensions diagram
Min 50
2-?7 /1
unit
Rack
Rack
144
160
±0.5
48
44
178
160
145.2
?4.5
135
165
121.2
3.8
Unit: mm
*
* Installation pitch between 2 rackmount
transmitters shall be more than 60mm for
environment where ambient temperature
exceeds 40degC.
5.8 Transmitter Installation
() Locate the meter sufficiently away from mechanical vibration or from corrosive gas atmosphere.
() Locate the meter where humidity is low and where temperature is as moderate as possible.
() Installation pitch between rackmount transmitters shall be more than 0mm for environment
where ambient temperature exceeds 0degC.
() Install in a location easy to access for inspection and maintenance.
() Install the transmitter with the dimensions shown in the figure below.
Fig..
Unit: mm