Signs of changes in or loss of functionality during use
Noticeable functional changes may include poor response, incomplete
extension, diminished swing phase control or stance phase stability, abnor
mal noises, etc.
4 Scope of Delivery
The scope of delivery for the product is shown on page 1 (cover picture).
The following single components and accessories are included in the scope
of delivery in the specified quantities and are available for reordering as
single components ( ), single components with minimum order quantities
( ) or single-component packs ( ):
Fig.
Item No. Quantity Designation
Reference number
–
–
1
Instructions for Use
647G813
–
–
1
Quick Reference Guide
647H537
–
–
–
1
Sport Knee Joint
–
–
–
1
Adjustment Wrench
710H10=2x3
–
–
2
Pyramid Dome Protector
Kit
2Z11=KIT
5 Preparation for Use
CAUTION
Incorrect alignment, assembly or adjustment
Injuries due to incorrectly installed or adjusted as well as damaged pros
thesis components
►
Observe the alignment, assembly and adjustment instructions.
CAUTION
Use of talcum
Falling, damage to the prosthetic knee joint due to lack of lubrication
►
Do not use talcum on the prosthetic knee joint or other prosthesis com
ponents.
►
Inform the patient.
The three-dimensional alignment of the prosthesis socket and the modular
components affects the function of the prosthesis. The patient can only
benefit from the special characteristics of the prosthetic knee joint when cor
rectly aligned.