15
CAUTION
Contact with heat, embers or fire
Risk of injury (such as burns) and risk of product damage.
►
Keep the product away from open flames, embers and other sources of heat.
NOTICE
Contact with oils, salves, lotions or other products that contain oils or acids
Insufficient stabilisation due to loss of material functionality
►
Do not expose the product to oils, salves, lotions or other products that contain oils or acids.
NOTICE
Exposure of the product to unsuitable environmental conditions
Damage, brittleness or destruction due to improper handling
►
Avoid storage in condensing ambient humidity.
►
Avoid contact with abrasive substances (e.g. sand, dust).
►
Do not expose the product to temperatures below -10 °C or above +50 °C (e.g. sauna,
excessive sunlight, drying on a radiator).
5 Handling
INFORMATION
►
The daily duration of use and period of application are generally determined by the physi
cian.
►
The initial fitting and application of the product must be carried out by qualified personnel.
►
Instruct the patient in the handling and care of the product.
►
Instruct the patient to see a physician immediately if any exceptional changes are noted (e.g.
worsening of the complaint).
5.1 Size Selection
►
Select the orthosis size based on the shoe size (see sizing table).
Size
Shoe size
Size
Heel container
Size
Longitudinal arch
piece
Size
Forefoot cap
S
36–41
Container size S
(36–41)
Longitudinal arch
piece size S (36–41)
Forefoot cap size S
(36–41)
M
42–46
Container size M–L
(42–50)
Longitudinal arch
piece size M (42–46)
Forefoot cap size M
(42–46)
L
47–50
Container size M–L
(42–50)
Longitudinal arch
piece size L (47–50)
Forefoot cap size L
(36–41)
5.2 Adaptation
1) Place the longitudinal arch piece (12) in the heel container (13).
INFORMATION: To determine the placement, affix the longitudinal arch piece to a
piece of paper. Position the patient on the longitudinal arch piece, align it with the
navicular bone (os naviculare) and mark the contour of the foot using a pencil (see
fig. 2). Then cut out the contour (see fig. 3) and place it in the heel container (see
fig. 4) so the end of the heel meets the mark in the heel container.
2) Mark the position of the longitudinal arch piece in the heel container (see fig. 5).
3) Fix the longitudinal arch piece along the mark inside the heel container.
Summary of Contents for 28F10
Page 2: ...1 2 3 4 5 2...
Page 3: ...6 7 8 9 10 11 3...
Page 4: ...12 13 14 15 16 17 4...
Page 5: ...18 19 20 21 22 23 5...
Page 6: ...24 Material PP PA PUR PU NBR SBR EVA PES PVC 6...
Page 87: ...87 3 3 3 3 1 3 3 2 3 4 Ilizarov 3 5 4 8 12 4 1 Hallgrimsson 6 1 8 2 2 10 3 11 5 6 12 4 4 1...
Page 88: ...88 4 2 10 C 50 C 5 5 1 S 36 41 S 36 41 S 36 41 S 36 41...
Page 90: ...90 16 22 17 23 18 24 5 3 1 2 21 3 22 4 23 6...
Page 91: ...91 pH Derma Clean 453H10 1 2 1 2 3 30 C 4 7 8 8 1 8 2 8 3 CE 2017 745 1 2020 05 25 28F10...
Page 93: ...93 3 3 1 3 2 3 3 3 3 1 3 3 2 3 4 3 5 4 8 12 4 1 6 1 8 2 2 10 3 11 5 6 12 BG Unfallklinik...
Page 94: ...94 4 4 1 4 2 10 C 50 C 5...
Page 96: ...96 8 8 7 9 4 5 13 14 10 1 15 11 2 16 12 19 13 14 6 20 15 21 16 22 17 23 18 24 5 3...
Page 97: ...97 1 2 21 3 22 4 23 6 pH Derma Clean 453H10 1 2 1 2 3 30 C 4 7 8...
Page 100: ...100 4 4 1 4 2 10 C 50 C 5 5 1 S 36 41 36 41 36 41 36 41...
Page 102: ...102 1 2 21 3 22 4 23 6 pH Derma Clean 453H10 1 2 1 2 3 30 C 4 7 8 8 1 20151111 6...
Page 103: ...103 8 2 8 3 CE 2017 745 CE...