24 | Ottobock
13E205
7AssemblyintoaProstheticComponent
Apply633F11SiliconeGreasetoallcontactbushingsoftheMyoRotronic,theredprotec-
tiveplugandtheconnectioncables.Thegreasewillprotecttheconnectionfromhumidity.
• Remove the protective cover from the motor of the Electric Wrist Rotator (Fig. 2, item 1).
• Cut the lock ring (Fig. 1, item 2) where it is notched and remove it.
• Remove the red protective plug (Fig. 2, item 3).
• Remove the protective cover (Fig. 1, item 1) from the MyoRotronic.
SlidetheMyoRotronicontotheElectricWristRotatorsothatthemarking1or3remainsfree
(asillustrated).
a)Settingfor2electrodes
(Programs 1, 2 and 3)
Inrightarmprostheses
3
3
CLOSE
OPEN
Motor
Battery
Switch
Buzzer
Electrode
Electrode
Inleftarmprostheses
1
1
CLOSE
OPEN
Motor
Battery
Switch
Buzzer
Electrode
Electrode
No. 1 remains free
No. 3 remains free
b)Settingfor1switch
(Program 4)
Inrightarmprostheses
3
3
CLOSE / PRO
OPEN / SUP
Motor
Battery
Switch
Electrode
Electrode
Inleftarmprostheses
1
1
CLOSE / PRO
OPEN / SUP
Motor
Battery
Switch
Electrode
Electrode
No. 1 remains free
No. 3 remains free