Ottobock | 21
13E205
A
Upperthreshold
B
Upperthreshold
Leftarm
prosthesis
Rotation: Pronation
Hand: CLOSE
MyoSoft/ PAULA:
blue line
Rotation: Supination
Hand: OPEN
MyoSoft/ PAULA:
red line
Rightarm
prosthesis
Rotation: Supination
Hand: OPEN
MyoSoft/ PAULA:
red line
Rotation: Pronation
Hand: CLOSE
MyoSoft/ PAULA:
blue line
OPEN/
CLOSE
PRO/SUP
Quickly beginning
muscle contraction
Thresholds A, B adjustable with MyoSelect
Slowly beginning
muscle contraction
Insert
battery
Program 2:
Cocontraction
Control with 2 electrodes
Switching between gripping function and rotation is performed
with Cocontraction. See section 5.2 "Switching". In addition, a
time control feature can be used to switch back to the gripping
function. The switch-back time is set using the 757T13 Myo-
Select. The respective setting can be read on scale C on the
MyoSelect's display.
Rotation can be controlled both digitally and proportionally (see
Section 5.1.2).
MyoSelect
scale C
Switch-
back
time
Insert
battery
Cocontraction or
automatically after
2-14 seconds
Cocontraction
OPEN/
CLOSE
PRO/SUP
0
2 sec.
1
4 sec.
2
6 sec.
3
8 sec.
4
10 sec.
5
12 sec.
5,9
14 sec.
6
Automatic switch-back deactivated; switch-back exclusively by cocontraction.
Upon delivery, the automatic switch-back is deactivated. The scale C in the MyoSelect's display
is therefore set to 6.