䊳
Set/read orifice rim heater mode (OTT Pluvio
2
S with orifice rim heater)
aOCHb!
ab<CR><LF>
Set mode
aOCH! ab<CR><LF>
Read mode
a
– sensor address
b
– 0 = Orifice rim heater disabled
1 = Mode #1; continuously ON
2 = Mode #2; temperature range control
3 = Modus 3; NWS Standard
4 = Modus 4; precipitation events
䊳
Set/read orifice rim heater target temperature (OTT Pluvio
2
S with orifice rim heater)
aOCHSpb!
apb<CR><LF>
Set target temperature
aOCHS!
apb<CR><LF>
Read target temperature
a
– sensor address
p
– sign (+)
b
– 2 … 9 (+2 … +9 °C);
factory s4 °C
䊳
Set/read orifice rim heater lower temperature limit (OTT Pluvio
2
S with orifice rim heater)
Orifice rim heater mode: #2
aOCHGpbb!
apbb<CR><LF>
Set lower temperature limit
aOCHG!
apbb<CR><LF>
Read lower temperature limit
a
– sensor address
p
– sign (+,–)
bb
– 40 … 9 (–40 … +9 °C);
factory setting –30 °C
䊳
Set/read orifice rim heater on-time/after-run time (OTT Pluvio
2
S with orifice rim heater)
Orifice rim heater mode: #3, #4
aOCHDbbbb!
abbbb<CR><LF>
Set on-time/after-run time
aOCHD!
abbbb<CR><LF>
Read on-time/after-run time
a
– sensor address
bbbb
– 1 … 1440 (1 … 1440 minutes);
factory setting 20 minutes
䊳
Set/read orifice rim heater starting time (OTT Pluvio
2
S with orifice rim heater)
Orifice rim heater mode: #3
aOCHZhh:mm:ss!
ahh:mm:ss<CR><LF>
Set starting time
aOCHZ!
ahh:mm:ss<CR><LF>
Read starting time
a
– sensor address
hh:mm:ss
– 00:00:00 … 23:59:59;
factory setting 14:00:00 Uhr
䊳
Set/read orifice rim heater self-test interval (OTT Pluvio
2
S with orifice rim heater)
Orifice rim heater mode: #1, #2, #3, #4
aOCHTbbbb!
abbbb<CR><LF>
Set self-test interval
aOCHT!
abbbb<CR><LF>
Read self-test interval
a
– sensor address
bbbb
– 1 … 10 080 (1 … 10 080 minutes);
factory setting 60 minutes*
* by OTT Pluvio
2
operating software adjustable values: 1, 12, 24 hours; 2, 3, 4, 5, 6, 7 days