36
17 Error messages
If erroneous measurements occur, or if an operating fault occurs, the OTT Orpheus
Mini stores one of the following error messages in the data memory instead of a
measured value.
Err 00
internal error (automated measurement)
Err 01
internal error (AD conversion error)
Err 03
Exceeded measuring range
Err 10
Measured value (still) not recorded
Internal errors indicate a device defect if they occur repeatedly.
18 Troubleshooting/fault correction
No communication possible
(operating program/OTT Orpheus Mini)
Password programmed?
➝
input correct password.
Batteries installed?
➝
install batteries.
Battery voltage lower than about 3.6 V?
➝
insert new batteries.
Battery contacts corroded?
➝
carefully clean the battery contacts.
Pressure probe cable damaged?
➝
send OTT Orpheus Mini to the factory for repair.
Infrared interface dirty?
➝
clean the infrared interface carefully with a damp, soft cloth.
Setting measured value not possible?
➝
scaling module missing. Check configuration*
(for advanced operation, see online help).
Communication starts and then breaks off
Battery voltage lower than about 3.6 V?
➝
insert new batteries.
Distance from readout unit/IrDA interface to OTT Orpheus Mini
too small/large?
➝
maintain correct distance (see Chapter 7.2).
Erroneous measured values
Temperature values erroneous
➝
check configuration* (for advanced operation, see online help).
Pressure sensor values erroneous
➝
check configuration* (for advanced operation, see online help).
➝
Check the pressure compensation capillary in the communication unit
(blocked?). If necessary, clean.
➝
Check pressure sensor for contamination. Clean if necessary.
No measured values in database
Configuration incorrect?
➝
check configuration* (for advanced operation, see online help).
Battery voltage lower than about 3.6 V?
➝
insert new batteries.
* The standard factory configuration is located on the "OTT Orpheus Mini Software" CD-ROM .