AURICAL OTOcam 300 is intended for use in the electromagnetic environment specified below. The user of AURICAL OTOcam 300 should ensure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
6 V rms
IISM Bands and Amateur
3 V rms
150 kHz to 80 MHz
6 V rms
IISM Bands and Amateur
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Proximity fields from RF wire-
less communications
IEC 61000-4-3
27 V/m
386 MHz
28 V/m
450 MHz,
9 V/m
710 MHz, 745 MHz, 780 MHz
28 V/m
810 MHz, 870 MHz, 930 MHz,
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
2450 MHz,
9 V/m
5240 MHz, 5500 MHz, 5785 MHz
27 V/m
386 MHz
28 V/m
450 MHz,
9 V/m
710 MHz, 745 MHz, 780 MHz
28 V/m
810 MHz, 870 MHz, 930 MHz,
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
2450 MHz,
9 V/m
5240 MHz, 5500 MHz, 5785 MHz
Separation distance between any electronic parts of
AURICAL OTOcam 300 and any RF wireless com-
munication equipment must be more than 30 cm (11.8
inches).
Note
: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
IEC 60601-1-2:2007 and EN 60601-1-2:2007
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
AURICAL OTOcam 300 is intended for use in the electromagnetic environment specified below. The user of AURICAL OTOcam 300 should ensure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
AURICAL OTOcam 300 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
AURICAL OTOcam 300 is suitable for use in all environments, including domestic environments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Not applicable
Otometrics - AURICAL OTOcam 300
29