CHAPTER TWENTY-THREE
Technical specifications
140
Technical specifications
Guidance and manufacturer’s declaration – electromagnetic emissions
The Otocheck ABR is intended for use in the electromagnetic environment
specified below. The user of the Otocheck ABR should ensure that it is used
in such an environment.
Emissions test
Compliance Electromagnetic environment
- guidance
RF emissions
Group 1
The Otocheck ABR uses RF energy
CISPR 11
only for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby equipment.
RF emissions
Class B
The Otocheck ABR is suitable
CISPR 11
for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
Harmonic emissions Class A
purposes.
IEC 61000-3-2
Voltage fluctuations/ Complies
flicker emissions
IEC 61000-3-3
23.2
Electromagnetic Compatibility:
Otocheck ABR
The Otocheck ABR should be put into service according to the EMC
(Electromagnetic Compatibility) information provided here.
Portable and mobile RF (Radio Frequency) communications equipment can affect
the operation of the Otocheck ABR. In particular, mobile telephones (‘cellphones’)
should not be operated within 10m of the Otocheck ABR.
The use of probes, electrodes, chargers and connection cables other than those
supplied by Otodynamics Ltd, and specifically for use with the Otocheck ABR may
result in increased emissions or decreased immunity of the Otocheck ABR.
The Otocheck ABR should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary the Otocheck ABR should be
observed to verify normal operation in the configuration in which it is used.