Equipment troubleshooting
Select
Reset
to reset the Otocheck to factory default settings. Changes
from the default setting and any users or worklist patients added to the
device will be lost. No test data will be removed.
Select
Format
to reformat the Otocheck database. Any records held on
the device will be irrecoverably lost.
The
Format
and
Reset
options are only available to users with Admin
rights.
18.4
Instrument fault message
In the event of an instrument fault, the following message will be displayed
at the start a test:
No stimulus will be delivered from the Otocheck probe and you will not be
able to start a test. Turn off the device and then switch it on again.
Important Note:
The
Instrument Fault
message can be triggered by a partially
connected probe. Ensure that the probe is fully connected and
the knurled sleeve screwed up correctly. (See
Connecting the
probe
in the
Getting started
section.
Run the probe checks (see chapter
13
). If the tests are ‘OK’ the device is
functioning correctly and can be used for testing again.
If you receive the
Instrument fault
message again, contact your dealer or
Otodynamics for support.