Otocheck ABR
User Manual
101
Quality checks
13.1.1
Results
The possible results of the test are:
Pass
The levels recorded at all frequencies are within the absolute range and
within ± 3dB of the probe specific values.
Fail
One or more of the levels recorded are outside the absolute range
specified for the probe. If a Fail is shown on screen inspect the probe
coupler tubes for debris, replace the coupler if necessary and repeat the
Probe Test, by selecting
Retest
, ensuring the ear piece is firmly inserted
in the test cavity. If the test continues to fail there may be a fault with the
probe or system. Contact your dealer or Otodynamics for advice.