background image

Specifications

Size: 

3.3 in H X 

2

.0 in W X 1.1 in D/ 85mm X 50 mm X 28mm

Weight: 

76 g / 2.7 oz

Humidity:

5% to 95% non-condensing relative humidity 

Alarms: 

Visual, vibrating, audible (minimum 95dB) 

Battery Life: 

24 months of operation/ 2 minutes of alarm per day.

Event Log Storage: 

Last 25 events. Newer events replace older events.

Certifications:

Class I, Groups A, B, C, D T4 ETL 4003191

Europe: 

 0359   II 1 G Ex ia IIC T4 Ga IECEx Ex ia IIC T4 Ga

ITS11ATEX27326 IECEx ETL 11.0002

-40°C<Tamb<50°C-CO

-40°C<Tamb<50°C-H2S

-35°C<Tamb<50°C-O2

Warranty:

2 years

12

Summary of Contents for OI-2100 TOCSIN

Page 1: ...Monitor ________________________________________ Operation Manual Revision 2 1w _______________________________________________________________________________________________ ________________________...

Page 2: ...tisseement 4 First Time Activation 5 Display Details 5 Day to Day Usage 6 Additional Steps 7 Failures FAQ s 8 Alarms 9 Alarm Types 9 Alarm Set Points 9 Event Log 10 Bump Test Interval bUP Clearing the...

Page 3: ...el OI 2100 TOCSIN Single Gas Personal Monitor This document should be read before initial operation of the product Should a question arise during the use of the product this document will serve as a f...

Page 4: ...ny obstructions i e debris or blockage Perform the Self Test to ensure proper operation of visual audible and vibrating alarms Confirm receipt of a check mark in the manual test is to be performed mak...

Page 5: ...TE The display mode can be changed in the IR Link software with the Sensor Reading and Life Remaining user options WARNING Users must familiarize themselves with the icons in both non alarm and alarm...

Page 6: ...beep all LEDs light up the unit vibrates all LCD display elements appear Screen 3 After the full element LCD screen the low alarm and high alarm set points will be displayed NOTE These alarm set point...

Page 7: ...software Screen 5 If the detector has been exposed to gas exceeding the low alarm set point a value will appear with max next to it This represents the peak value highest that the detector has seen Af...

Page 8: ...the LCD will display EO5 In case of an EO5 screen discontinue use and contact Otis Instruments for a replacement detector Along with the battery the sensor is continuously monitored during normal oper...

Page 9: ...hen the unit has eight hours remaining it will begin to beep flash and vibrate intermittently NOTE To end the alarms press the button down Once the unit has reached the end of its operating life the d...

Page 10: ...re from 1 to 365 days NOTE The unit default is to have no bump interval programmed If a detector is due for a bump test the unit display will alternate between the months remaining and bUP In addition...

Page 11: ...ay Detail O2 CAL NOTE Only perform O2 calibration in normal ail 20 9 Oxygen that is free of hazardous gases 1 Press and hold the yellow button for four seconds 2 The display screen will show CAL and t...

Page 12: ...dible minimum 95dB Battery Life 24 months of operation 2 minutes of alarm per day Event Log Storage Last 25 events Newer events replace older events Certifications Class I Groups A B C D T4 ETL 400319...

Page 13: ...Proper storage installation calibration use maintenance and compliance with the user s manual instructions and any other applicable recommendations of Otis Instruments 2 The buyer promptly notifying...

Page 14: ...Otis Instruments Inc 2200 E Villa Maria Dr Bryan TX 77802 Service Department 903 566 1300 Corporate Office 979 776 7700 service otisinstruments com www otisinstruments com 14...

Reviews: