Event Log
By default the detector stores the last twenty five (25) alarm events. The system stores events by first in first out, i.e. the
26
th
event will replace the first event and so on. This information can be downloaded using the OI-TOCSIN-LINK. For
each alarm event the detector records the following:
•
The detector serial number
•
Bump Test (Yes or No)
•
Life remaining on the detector
•
Number of self tests performed
•
Number of events
•
Alarm Condition (Low, High, or OL)
•
Specific event date and time
•
Peak gas concentration in ppm or %.
“Bump Test Interval” (bUP) - Clearing the alert
Using the IR Link or Software Manager, units can be programmed to alert the user if a bump test is due. This interval
can be set anywhere from 1 to 365 days.
NOTE: The unit default is to have no bump interval programmed.
If a detector is due for a bump test, the unit display will alternate between the months remaining and “bUP”. In addition
the unit will emit alternating flashes (left and right) every 5 seconds. And the “test” icon will remain even after a button
push.
This alert can be cleared by either placing the unit in an OI-2100-DOCK or by manually applying gas to the unit. To
manually clear the alert, press the button down once, the words "GAS" will show on the display while the TEST icon
flashes. The detector will wait for 45 seconds for the target gas to be applied, or a button press to skip the bump test. If
the SGC is bumped while showing "GAS", then it will record as a bump test in the event log instead of as an exposure.
If no gas is applied, it will return to the normal screen and will not record anything in the event log.