OTIS OI-2100 TOCSIN Operation Manual Download Page 10

Event Log

By default the detector stores the last twenty five (25) alarm events. The system stores events by first in first out, i.e. the 
26

th 

event will replace the first event and so on. This information can be downloaded using the OI-TOCSIN-LINK. For 

each alarm event the detector records the following:

The detector serial number

Bump Test (Yes or No)

Life remaining on the detector

Number of self tests performed

Number of events

Alarm Condition (Low, High, or OL)

Specific event date and time

Peak gas concentration in ppm or %.

“Bump Test Interval” (bUP) - Clearing the alert

Using the IR Link or Software Manager, units can be programmed to alert the user if a bump test is due. This interval 
can be set anywhere from 1 to 365 days.

NOTE: The unit default is to have no bump interval programmed.

If a detector is due for a bump test, the unit display will alternate between the months remaining and “bUP”. In addition 
the unit will emit alternating flashes (left and right) every 5 seconds. And the “test” icon will remain even after a button  
push.

This alert can be cleared by either placing the unit in an OI-2100-DOCK or by manually applying gas to the unit. To 
manually clear the alert, press the button down once, the words "GAS" will show on the display while the TEST icon 
flashes. The detector will wait for 45 seconds for the target gas to be applied, or a button press to skip the bump test. If 
the SGC is bumped while showing "GAS", then it will record as a bump test in the event log instead of as an exposure. 
If no gas is applied, it will return to the normal screen and will not record anything in the event log.

10

Summary of Contents for OI-2100 TOCSIN

Page 1: ...Monitor ________________________________________ Operation Manual Revision 2 1w _______________________________________________________________________________________________ ________________________...

Page 2: ...tisseement 4 First Time Activation 5 Display Details 5 Day to Day Usage 6 Additional Steps 7 Failures FAQ s 8 Alarms 9 Alarm Types 9 Alarm Set Points 9 Event Log 10 Bump Test Interval bUP Clearing the...

Page 3: ...el OI 2100 TOCSIN Single Gas Personal Monitor This document should be read before initial operation of the product Should a question arise during the use of the product this document will serve as a f...

Page 4: ...ny obstructions i e debris or blockage Perform the Self Test to ensure proper operation of visual audible and vibrating alarms Confirm receipt of a check mark in the manual test is to be performed mak...

Page 5: ...TE The display mode can be changed in the IR Link software with the Sensor Reading and Life Remaining user options WARNING Users must familiarize themselves with the icons in both non alarm and alarm...

Page 6: ...beep all LEDs light up the unit vibrates all LCD display elements appear Screen 3 After the full element LCD screen the low alarm and high alarm set points will be displayed NOTE These alarm set point...

Page 7: ...software Screen 5 If the detector has been exposed to gas exceeding the low alarm set point a value will appear with max next to it This represents the peak value highest that the detector has seen Af...

Page 8: ...the LCD will display EO5 In case of an EO5 screen discontinue use and contact Otis Instruments for a replacement detector Along with the battery the sensor is continuously monitored during normal oper...

Page 9: ...hen the unit has eight hours remaining it will begin to beep flash and vibrate intermittently NOTE To end the alarms press the button down Once the unit has reached the end of its operating life the d...

Page 10: ...re from 1 to 365 days NOTE The unit default is to have no bump interval programmed If a detector is due for a bump test the unit display will alternate between the months remaining and bUP In addition...

Page 11: ...ay Detail O2 CAL NOTE Only perform O2 calibration in normal ail 20 9 Oxygen that is free of hazardous gases 1 Press and hold the yellow button for four seconds 2 The display screen will show CAL and t...

Page 12: ...dible minimum 95dB Battery Life 24 months of operation 2 minutes of alarm per day Event Log Storage Last 25 events Newer events replace older events Certifications Class I Groups A B C D T4 ETL 400319...

Page 13: ...Proper storage installation calibration use maintenance and compliance with the user s manual instructions and any other applicable recommendations of Otis Instruments 2 The buyer promptly notifying...

Page 14: ...Otis Instruments Inc 2200 E Villa Maria Dr Bryan TX 77802 Service Department 903 566 1300 Corporate Office 979 776 7700 service otisinstruments com www otisinstruments com 14...

Reviews: