![background image](http://html1.mh-extra.com/html/otis/gen2-notis-quad-oi-6940/gen2-notis-quad-oi-6940_operation-manual_1658589029.webp)
Sensor Last Time Null/Cal
1. Touch
ADD
(NEXT) to continue to the next setting: Sensor 1 Last Time Null/Cal.
2. Touch
MENU
(SELECT) to view the last time Sensor 1 was nulled and calibrated; touch
ADD
(NEXT) to continue
to the next menu option.
3. If
MENU
was touched, the Display Screen should resemble the following illustration:
NOTE: If the sensor has never been nulled or
calibrated, or if the null/cal values have been reset
since the last time the sensor was nulled/caled, the
Display Screen will show “NEVER”.
NOTE: The Last Time Since Null/Cal is only
calculated when the Notis Quad is turned On.
NOTE: The Last TimeSince Null/Cal is only saved
in hours and days. Therefore, if the last time the
sensor was nulled/caled was 59 minutes and 59
seconds ago (or if the unit was turned off 59 minutes
and 59 seconds after the null/cal was complete), the
Last Time Since Null/Cal reading would be
“NEVER” (because an hour had not been reached).
4. Once the Sensor 1 Last Time Null/Cal information has been viewed, touch
MENU
(BACK) or
BACK
(BACK).
29