19
Figure 8.9. Run screen
Figure 8.10. ‘TRAY OPEN’ warning.
8.5 Operational Safety
(I)
Safety Interlock
If the tray is opened while the system is running the program will automatically stop. Turn the lamp OFF and
reset the
Set Time to 00:00. The display will indicate ‘TRAY OPEN’ (as shown in Figure 8.10).
(II)
High Temperature Safety Warning
If one or more of the internal temperature sensors reads above certain temperature value, a high
temperature warning will appear on the screen.
Table 8.2
shows the different warnings for this unit for given
temperatures. The error code that is displayed when the tray or electronics sensor exceeds 65 °C indicates
which sensor is at a high temperature and is used for troubleshooting.
(III)
Temperature Error
If one of the temperature sensors gives an error reading, a warning (as shown in
Figure 8.11) will appear and
the user is required to restart the unit. The error code that is displayed indicates which sensor is in error and
is used for troubleshooting.