background image

19

Figure 8.9. Run screen

Figure 8.10. ‘TRAY OPEN’ warning.

8.5 Operational Safety

(I) 

Safety Interlock

If the tray is opened while the system is running the program will automatically stop. Turn the lamp OFF and 
reset the 

Set Time to 00:00. The display will indicate ‘TRAY OPEN’ (as shown in Figure 8.10).

(II) 

High Temperature Safety Warning

If one or more of the internal temperature sensors reads above certain temperature value, a high 
temperature warning will appear on the screen. 

Table 8.2

 shows the different warnings for this unit for given 

temperatures. The error code that is displayed when the tray or electronics sensor exceeds 65 °C indicates 
which sensor is at a high temperature and is used for troubleshooting.

(III) 

Temperature Error

If one of the temperature sensors gives an error reading, a warning (as shown in 

Figure 8.11) will appear and 

the user is required to restart the unit. The error code that is displayed indicates which sensor is in error and 
is used for troubleshooting. 

Summary of Contents for UV Ozone Cleaner

Page 1: ...USERMANUAL Manual Version 2 0 E Product Code L2002A2 Product Version 2 0 Software Version 2 0 UV OZONE CLEANER ...

Page 2: ...fety Operation 10 3 9HealthandSafety Servicing 11 4 Unpacking 11 4 1PackingList 11 4 2DamageInspection 11 5 Specifications 12 6 SystemComponents 12 7 Installation 13 8 Operation 14 8 1Overview 14 8 2UserInterface 15 8 3PracticalOperation 15 8 4ProgramOperation 16 8 5OperationalSafety 19 9 Maintenance 20 9 1Cleaning 20 9 2RepairandService 21 9 3StorageConditions 21 10 Troubleshooting 21 11 Relatedp...

Page 3: ...lyofthe orderof0 1 1 0Å s Theexactexposuretimesandmethodsrequiredtoremovevariouscontaminantswill varyandshouldbedeterminedempirically TheOssilaSolarCellPrototypingPlatformisacomplementarycollectionofsubstrates materials and equipmentaspartofahigh performancestandardphotovoltaicreferencearchitecture Thisplatform enablesresearcherstoproducehigh quality fully functionalsolarcellswhichcanbeusedasareli...

Page 4: ...Cleaner L2002A2 Serialnumber L2002A 2000 2000 2000 xxxx Objectofdeclaration UVOzoneCleaner L2002A2 Theobjectofdeclarationdescribedaboveisinconformity withtherelevantUnionharmonisationlegislation LowVoltageDirective2014 35 EU EMCDirective2014 30 EU RoHSDirective2011 65 EU Thefollowingharmonisedstandardsandtechnicalspecifications havebeenapplied BSEN61010 1 2010Safetyrequirementsforelectricalequipme...

Page 5: ...τότηταςΕΕ Κατασκευαστής OssilaLtd SolproBusinessPark WindsorStreet S47WD ΗνωμένοΒασίλειο ΔηλώνουμευπεύθυναόnοαναφερόμενοςεξοπλισμόςσυμμορφώνεταιμετησχεnκήνομοθεσίαεναρμόνισηςτηςΕΕπουυπάρχει σnςπροηγούμενεςσελίδεςτουπαρόντοςεγγράφου Español DeclaracióndeconformidadUE Fabricante OssilaLtd SolproBusinessPark WindsorStreet S47WD ReinoUnido Declaramosbajonuestraúnicaresponsabilidadqueelsiguienteproduct...

Page 6: ...dsorStreet S47WD ReinoUnido Declarasobsuaexclusivaresponsabilidadequeoequipamentoindicadoestáemconformidadecomalegislaçãodeharmonização relevantedaUEmencionadana s página s anterior es destedocumento Română DeclaraţiedeconformitateUE Producător OssilaLtd SolproBusinessPark WindsorStreet S47WD RegatulUnit DeclarăpeproprierăspunderecăechipamentulprezentatesteînconformitatecuprevederilelegislaţieiUEd...

Page 7: ...edinthefollowingenvironmental conditions Indoorsinalaboratoryenvironment pollutiondegree2 Altitudesupto2000m Temperaturesof5 Cto40 C maximumrelativehumidityof80 upto31 C Thecleanerissuppliedwithapowercord andifnecessaryanadditionalpowertransformer forthecountry ofpurchase inaccordancewithEuropeanCommissionregulationsandBritishStandards Useofanyother electricalpowercables adaptors ortransformersisn...

Page 8: ...nnectthepowercordandwait10minutes highvoltagemaypersist incapacitorsforsometimeafterremovalofpower III CAUTION TheUVOzoneCleanerusesaground typepowerplug whichmustbe connectedtoagroundedoutlettopreventelectricalshock TheUVOzoneCleaning unitwillbesuppliedwithanearthedplugappropriateforthecountryofpurchase IV TheUVlampisamercuryvapourlamp Theusermaybeexposedtomercuryif thelampglassbreaks Ifthishappe...

Page 9: ...atingequipment II TheUVOzoneCleanerisintendedforoperationinsideafully functioningfumehood orfumecupboardonly ThereisnoozonefiltrationsystemincorporatedintotheUV OzoneCleaner I UVradiationcancausesevereburnstotheeyesandskin Anintegratedsafety interlockpreventsthelampfrombeingpoweredonwhilstthetrayisopen Asa precaution neverlookdirectlyintothesamplechamberwhentheUVozonecleaner unitispoweredon II Ozo...

Page 10: ...ofanypurchased equipment Ifchangesaremade suchalterationsmaydamagetheequipment causeinjury orevendeath Attheveryleast suchalterationswillvoidyourwarranty ThestandarditemsincludedwiththeUVOzoneCleanerare UVOzoneCleanerunit Powercordset specificforcountryofoperation Usermanual 4 Unpacking 4 2DamageInspection Examinethecomponentsforevidenceofshippingdamage Ifdamagehasoccurred pleasecontactOssila dire...

Page 11: ...owninTable5 1 Table5 1 UVOzoneCleanerspecifications 6 SystemComponents UVOzoneCleaner Specifications UVLamp Syntheticquartz4x4gridUVlamp UVlampkeywavelengths 185nmand254nm Powersupply 230V 10 50 60Hz 50VA Classofprotection ClassI Degreeofprotection IP20 Maximumruntime 59minutesand59seconds Safetyfeatures Interlock thermalcut out Unitdimensions Width 193mm Height 230mm Depth 300mm Traydimensions 10...

Page 12: ...afumecupboard i Ensuretheareaisfreefromvibrations temperatureextremesandhighlyflammableor explosivematerials 2 BeforepluggingintheUVOzoneCleaner ensurethepowerswitchontheunitisswitchedtothe 0 position off 3 ConnectthepowercordtotheUVOzoneCleanerunit i SeeFigure7 1toseehowtoconnecttheUVOzoneCleanertothepowersupplycable 4 SwitchtheUVOzoneCleanerpowerswitchtothe I positiontoturnon ...

Page 13: ...heUVOzoneCleanerpowercordcable Figure8 1 ComponentsoftheUVOzoneCleaner Atop downviewoftheUVOzoneCleanerisshowninFigure8 1 withalltherelevantcomponents highlighted Figure8 2showsthepowercordsocket fusesocket andthepowerswitch Durablemetalcasing Simplekeypadforprogramming Substratetray ClearLCDdisplay Internalsafetyinterlockpreventsthe lampbeingonwhentrayisopen ...

Page 14: ...programandresetstheSetTimetozero MIN Increasethesettimeby1minute Decreasethesettimeby1minute SEC Increasethesettimeby10seconds Decreasethesettimeby10seconds Increasethesettimeby1second Decreasethesettimeby1second 8 2UserInterface 8 3PracticalOperation Operatewithinfumecupboardinaroomwithsuitableairventilationatalltimes 1 Openthetraydoor 2 Carefullyloadyoursampletobecleanedontothetray I Thetraysurf...

Page 15: ...erateatamaximumof4000V Makesuretheunitis connectedtoagroundedpoweroutlet Ensuretheareaaroundthebackoftheunit iskeptclear Itisgoodpracticetokeepanycablesaroundtheunitinatidymanner Avoidkeepinganysolventsaroundthispieceofequipment WARNING Ultravioletlight Thelow pressuremercuryvapourgridlampinsidethisequipmentemitsharmfulUV radiation Avoidexposureatalltimes Avoidkeepinganysolventsaroundthispiece ofe...

Page 16: ...RTbuttontoproceedtocontinue 4 TheusermustreadthecautionmessageandpressSTARTtoenterthemainscreenoftheUVOzone CleanerasshowninFigure8 7 II Pressandholdbothdouble incrementsecondbuttons atthesametimeto enterthebuzzersettingsasinFigure8 6 III Tochangethebuzzersettings pressthesingle incrementupsecondbutton toturn thebuzzersoundon orthesingle incrementdownsecondbutton toturnthebuzzer soundoff Figure8 5...

Page 17: ...thetimerhasbeenset presstheSTARTbuttontoruntheprogram TheLCDwilldisplaythe followinginformation asshowninFigure8 9 I TheElapsedtimerwillbeginandtheLEDindicatorlightwillswitchon II WhentheElapsedtimematchestheSetTime theone notebuzzerwillsound indicatingthe programhasfinished III Whiletheprogramisrunning theusercanstoptheprogrambypressingtheSTOPbutton This willalsoresettheSetTimeto00 00 ...

Page 18: ...resensorsreadsabovecertaintemperaturevalue ahigh temperaturewarningwillappearonthescreen Table8 2showsthedifferentwarningsforthisunitforgiven temperatures Theerrorcodethatisdisplayedwhenthetrayorelectronicssensorexceeds65 Cindicates whichsensorisatahightemperatureandisusedfortroubleshooting III TemperatureError Ifoneofthetemperaturesensorsgivesanerrorreading awarning asshowninFigure8 11 willappear...

Page 19: ...removeanyoil grease orgrime Neveruseliquidsolventsordetergents Repairsorservicingnot coveredinthismanualshouldonlybeperformedbyqualifiedpersonnel 9 1Cleaning Figure8 11 Temperatureerrorwarning Table8 2 Hightemperaturewarning Warning Display HIGHTEMP iftrayorelectronicsis 50 C VERYHIGHTEMP iftrayorelectronicsis 65 C ...

Page 20: ...ctionandensurethepoweris turnedON b Ensuretheunitisfirmlypluggedintothepower supplyandtheplugisfirmlyconnectedtoboth thepowerinletandthepowersocket c Ensuretheunitisunplugged Checkthefuse ontherearpanel Ifithasblown replacewitha suitablyrated1Aslowblowfuse d ContactOssilaforareplacementtransformer e Ifalltheabovecauseshavebeenconsidered theremaybeafaultontheboard Pleasecontact Ossilaforinformation...

Page 21: ... syringe pumps for the automatic dispensing of solutions Source voltage measure current get data Simplify and acclerate your data collection Our range of 15 x 20 mm ITO substrates for OPV OLED and sensing applications Silicon substrates with thermal oxide layer pre cut to fit the Ossila OFET fabrication systems Produce high quality coatings without any substrate warping Perfect for busy labs with ...

Page 22: ......