3.9 Health and Safety – Servicing
4.1 Packing List
III. When operating the UV Ozone Cleaner, air flow in and around the unit must remain
unobstructed to prevent the unit from overheating. If the temperature inside the
sample chamber reaches above 65 °C, the unit will automatically switch off to
prevent overheating. Allow the system to cool down to room temperature before
resuming use.
IV. IMPORTANT NOTE: Excessive cooling will diminish the lamp’s operating voltage
and effectiveness, whereas Inadequate cooling may cause damage to system
components.
V. The UV Ozone Cleaner has a high voltage source. Do not attempt to open the unit.
I.
Service or installation work that includes integrating electrical components should
only be performed by an Ossila engineer. Never alter the wiring of any purchased
equipment. If changes are made, such alterations may damage the equipment,
cause injury, or even death. At the very least, such alterations will void your warranty.
The standard items included with the UV Ozone Cleaner are:
•
UV Ozone Cleaner unit
•
Power cord set (specific for country of operation)
•
User manual
4. Unpacking
4.2 Damage Inspection
Examine the components for evidence of shipping damage. If damage has occurred, please contact Ossila
directly for further action. The UV lamp is a mercury vapour lamp and the user may be exposed to mercury
only if the glass of the lamp is cracked or broken. If this happens, please contact Ossila after abiding by these
rules to minimise exposure:
1. Ensure the power is disconnected to avoid the risk of electrocution.
2. Store the container and UV Ozone Cleaner in a fume hood and leave the room under ventilation for at
least 15 minutes.
3. Using cut-resistant gloves, collect any stray pieces of glass in a sealed container. Dispose of the glass at a
suitable collection point.